Drugs & medicalClass II

Philips North America Llc recalls Philips DigitalDiagnost C50 -intended for use in generating radiographic images of human anatomy Model Number : 712201

If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through…

Read the official notice →
Date
2023-04-19
Country
United States
Brand
Philips
Product
Philips DigitalDiagnost C50 -intended for use in generating radiographic images of human anatomy Model Number : 712201
Classification
Class II
Units
Nationwide including Puerto Rico and Virgin Islands Foreign: ARGENTINA…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Philips recalls

  1. 2023-05-24
  2. US
    Drugs & medicalClass II

    Philips North America recalls Philips Incisive CT -Computed Tomography X-Ray System Model Number: 72813

    Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator's foot and injury

    FDA · DevicePhilips North AmericaWorldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN,…Official notice →
  3. US
    Drugs & medicalClass II

    Philips North America Llc recalls Spectral CT 7500 Computed Tomography X-Ray System, Model Number 728333

    Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operators foot. If foot or leg entrapment occurs while pressing Unload Pedal, the risk to operators may include:…

    FDA · DevicePhilips North America LlcWorldwide - US Nationwide distribution including in the states of States: AZ,…Official notice →
  4. 2023-05-10
  5. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips ProxiDiagnost N90 1.0 (Model 706100) and ProxiDiagnost N90 1.1 (Model 706110)- A multi-functional general radiography and…

    Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop…

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide including Puerto Rico and the countries…Official notice →
  6. 2023-04-19
  7. US
    Drugs & medicalClass II

    Philips North America Llc recalls CombiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 709030 and 709031

    If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell…

    FDA · DevicePhilips North America LlcUS Nationwide. Global Distribution.Official notice →
  8. 2023-01-18
  9. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips Fetal Spiral Electrode-intended for patients requiring fetal heart rate monitoring during labor. PN: 989803137631

    Fetal Spiral Electrode (FSE) spiral tip of the FSE breaking off during use and requiring surgical intervention to remove the broken tip from the neonate patient. Philips has decided to discontinue the distribution of…

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  10. 2022-12-14
  11. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips SmartPath to Ingenia Elition X MR System, Model Numbers 782118, 782132, 782144

    Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.

    FDA · DevicePhilips North America LlcUS Nationwide.Official notice →
All 100 Philips recalls →

Other drugs & medical recalls in the United States

  1. 2023-04-19
  2. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls stryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 6 RM/LL), REF 170616, intended for use in a Partial Knee Arthroplasty surgical procedure

    Certain lots of the Restoris MultiCompartmental Knee (MCK) Tibial Baseplate Trials contain product with incorrect size laser marking

    FDA · DeviceHowmedica Osteonics Corp.U.S., Netherlands, Japan, China, India, Korea, and Australia.Official notice →
  3. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls stryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee Arthroplasty surgical procedure

    Ccertain lots of the Restoris MultiCompartmental Knee (MCK) Tibial Baseplate Trials contain product with incorrect size laser marking

    FDA · DeviceHowmedica Osteonics Corp.U.S., Netherlands, Japan, China, India, Korea, and Australia.Official notice →
  4. US
    Drugs & medicalClass II

    Encore Medical, LP recalls EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104; EMPOWR Knee Patella Drill, REF: 802-05-108

    Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect…

    FDA · DeviceEncore Medical, LPUS: TX, IN, MD, CA, KS, AL, HI, WAOfficial notice →
  5. US
    Drugs & medicalClass II

    Angiodynamics, Inc. recalls NanoKnife 5-Probe Procedure Pack 15 cm - Activation. Indicated for the surgical ablation of soft tissue. Catalog Number: 20400111 UPN:…

    Not programmed in accordance with specification. The programming affects the RFID function and does not allow the NanoKnife probes to be recognized by the NanoKnife generator resulting in a delay in procedure

    FDA · DeviceAngiodynamics, Inc.US Nationwide Distribution: CA,CO, FL, IA, NY, OHOfficial notice →
  6. US
    Drugs & medicalClass II

    Edwards Lifesciences, LLC recalls Edward Lifesciences, REF 774F75, Swan-Ganz CCOmbo V, CCO/SvO2/CEDV/VIP, Thermodilution Catheter, Exterior Diameter 7.5F (2.5mm), Usable…

    There is a potential that the distal end of the catheter may be bent beyond its intended curvature and when connected to patient monitoring device, the monitor may display an error message.

    FDA · DeviceEdwards Lifesciences, LLCUS: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO…Official notice →
  7. US
    Drugs & medicalClass II

    Bio-Rad Laboratories, Inc. recalls BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD

    Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots.…

    FDA · DeviceBio-Rad Laboratories, Inc.Worldwide distribution - US Nationwide and the countries of Canada, Czech…Official notice →
See all drugs & medical recalls →