Bio-Rad Laboratories, Inc. recalls BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD
Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots.…
Read the official notice →- Date
- 2023-04-19
- Country
- United States
- Brand
- Bio-Rad Laboratories
- Product
- BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Canada, Czech…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Bio-Rad Laboratories recalls
- 2025-09-10
US
Bio-Rad Laboratories, Inc. recalls Platelia Toxo IgM Catalog Number 26211 UDI Code: 03610520005552 The Platelia Toxo IgM assay is a semi-quantitative immunocapture…
Due to a risk of false positive results that could lead to unnecessary medical treatment.
FDA · DeviceBio-Rad Laboratories, Inc.U.S. Nationwide distribution in the states of CA, CO, FL, NC and NV.Official notice →- 2024-04-03
US
Bio-Rad Laboratories recalls Bio-Rad, REF: 12005660, QXDx BCR-ABL %IS Kit
Due to incorrect calibrations and controls, their is a potential of incorrect reporting of lower %IS and molecular response values.
FDA · DeviceBio-Rad LaboratoriesUS Nationwide distribution in the states of PA, SD, TX, FL, MN.Official notice →- 2023-05-17
US
Bio-Rad Laboratories, Inc. recalls BioPlex 2200 REF 665-2050, APLS IgM Pack, APLS IgM Reagent Pack
APLS IgM reagent kits were packaged with the incorrect conjugate, which could lead to an increase in false-positive and false-negative results.
FDA · DeviceBio-Rad Laboratories, Inc.US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, FL,…Official notice →- 2022-12-07
US
Bio-Rad Laboratories, Inc. recalls CMV IgM EIA, in vitro diagnostic.
Due to an unusual increase in the positivity rate with human IgM antibodies to cytomegalovirus (CMV) Enzyme Immunoassay (EIA).
FDA · DeviceBio-Rad Laboratories, Inc.U.S.: CA, CO, FL, GA, HI, IA, MA, MN, NC, NH, NJ, PA, TN, and TX O.U.S.: NoneOfficial notice →- 2022-06-08
US
Bio-Rad Laboratories, Inc. recalls Resin UPDATE CD-ROM, REF: 250-3020 inside VARIANTnbs Sickle Cell Program Reorder Pack, REF: 250-3000
The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM software included in…
FDA · DeviceBio-Rad Laboratories, Inc.US: AZ, AR, GA, CO, TN, LA, VA, FL, KY, IN, KS, MD, MN, MO, NC, NJ, NY, OH,…Official notice →- 2022-05-25
US
Bio-Rad Laboratories, Inc. recalls BioPlex 2200 REF 12000670 ToRC IgM Reagent Pack
Performance concerns for the CMV IgM analyte. Analyte has demonstrated variable elevated background depending on the handling of the reagent pack.
FDA · DeviceBio-Rad Laboratories, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AR, AZ,…Official notice →
Other drugs & medical recalls in the United States
- 2023-04-19
US
Philips North America Llc recalls Philips DigitalDiagnost C50 -intended for use in generating radiographic images of human anatomy Model Number : 712201
If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through…
FDA · DevicePhilips North America LlcNationwide including Puerto Rico and Virgin Islands Foreign: ARGENTINA…Official notice →US
Angiodynamics, Inc. recalls NanoKnife 5-Probe Procedure Pack 15 cm - Activation. Indicated for the surgical ablation of soft tissue. Catalog Number: 20400111 UPN:…
Not programmed in accordance with specification. The programming affects the RFID function and does not allow the NanoKnife probes to be recognized by the NanoKnife generator resulting in a delay in procedure
US
Edwards Lifesciences, LLC recalls Edward Lifesciences, REF 774F75, Swan-Ganz CCOmbo V, CCO/SvO2/CEDV/VIP, Thermodilution Catheter, Exterior Diameter 7.5F (2.5mm), Usable…
There is a potential that the distal end of the catheter may be bent beyond its intended curvature and when connected to patient monitoring device, the monitor may display an error message.
FDA · DeviceEdwards Lifesciences, LLCUS: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO…Official notice →US
Western/Scott Fetzer Company recalls Digital VIPR ROC-9839 Roberts Oxygen Ready to Use Digital VIPR (head only) ROC-9840, Roberts Oxygen Ready to Use Digital VIPR…
Gauge malfunction: Digital gauge displays an incorrect tank pressure - displays pressure higher than actual tank pressure. This may lead to tank running out of oxygen during use without audible alarms, potentially…
FDA · DeviceWestern/Scott Fetzer CompanyUS Nationwide distribution in the states of CO, CT, ID, MD, MT, TN, WA.Official notice →US
Coloplast Manufacturing US, LLC recalls Titan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
XXX
FDA · DeviceColoplast Manufacturing US, LLCWorldwide distribution - US Nationwide and the countries of Turkey, Italy,…Official notice →US
Coloplast Manufacturing US, LLC recalls Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile…
FDA · DeviceColoplast Manufacturing US, LLCWorldwide distribution - US Nationwide and the countries of Turkey, Italy,…Official notice →