Dexcom, Inc. recalls Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system
A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are…
Read the official notice →- Date
- 2025-06-11
- Country
- United States
- Brand
- Dexcom
- Product
- Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system
- Classification
- Class II
- Units
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of:…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Dexcom recalls
- 2026-07-01
US
Dexcom, Inc. recalls Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog…
Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC.…
FDA · DeviceDexcom, Inc.Dexcom does not have distribution records for this product due to its…Official notice →- 2026-05-20
US
Dexcom, Inc. recalls Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software…
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV…
FDA · DeviceDexcom, Inc.SW12300 & SW13355 Dexcom G7 iOS and watchOS app versions 2.9.0 through 2.11.2:…Official notice →- 2025-11-05
US
Dexcom, Inc. recalls Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number:…
A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could…
FDA · DeviceDexcom, Inc.This product was distributed throughout the US. G6 Android US v1.15.0 SW11678…Official notice →- 2025-09-10
US
Dexcom, Inc. recalls Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software…
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App…
FDA · DeviceDexcom, Inc.SW12299 G7 Android CGM App Worldwide Distribution - US Nationwide and the…Official notice →- 2025-06-25
US
Dexcom, Inc. recalls Dexcom One+ Continuous Glucose Monitoring System, SKU(Part Number): STK-D7-013 (MT28285-1), STK-D7-014 (MT28285-1), STK-D7-103 (MT28285-2)
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low…
FDA · DeviceDexcom, Inc.worldwide distribution - US Nationwide and the countries of Andorra, Argentina…Official notice →- 2025-02-26
US
Dexcom, Inc. recalls Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OR-001
Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net…
FDA · DeviceDexcom, Inc.Worldwide - US Nationwide distribution in the states of OH, TX, PA, MN, SC,…Official notice →
Other drugs & medical recalls in the United States
- 2025-06-11
US
Healthmark Industries Co., Inc. recalls ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample)
During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries…
FDA · DeviceHealthmark Industries Co., Inc.US Nationwide and Internationally to countries of: Canada, Malaysia.Official notice →US
Beckman Coulter Mishima K.K. recalls Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 -…
Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of…
FDA · DeviceBeckman Coulter Mishima K.K.Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX,…Official notice →US
Beckman Coulter Mishima K.K. recalls Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 -…
A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing.…
FDA · DeviceBeckman Coulter Mishima K.K.Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX,…Official notice →US
VIRTUAL INCISION CORPORATION recalls MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version…
The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the…
US
Luminex Corporation recalls VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System;…
It was determined that the consumable issue may result in No Call - INT CTL 1 and/or False Negative results on a small portion of the lot of test cartridges.
FDA · DeviceLuminex CorporationUS Nationwide distribution in the states of AL, CA, MN, TX, AZ, NJ, WA, SC,…Official notice →US