Dexcom recalls
7 official recalls and safety alerts involving Dexcom since 2025-02, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Dexcom or the relevant agency.
- 2026-07-01
US
Dexcom, Inc. recalls Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog…
Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC.…
FDA · DeviceDexcom, Inc.Dexcom does not have distribution records for this product due to its…Official notice →- 2026-05-20
US
Dexcom, Inc. recalls Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software…
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV…
FDA · DeviceDexcom, Inc.SW12300 & SW13355 Dexcom G7 iOS and watchOS app versions 2.9.0 through 2.11.2:…Official notice →- 2025-11-05
US
Dexcom, Inc. recalls Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number:…
A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could…
FDA · DeviceDexcom, Inc.This product was distributed throughout the US. G6 Android US v1.15.0 SW11678…Official notice →- 2025-09-10
US
Dexcom, Inc. recalls Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software…
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App…
FDA · DeviceDexcom, Inc.SW12299 G7 Android CGM App Worldwide Distribution - US Nationwide and the…Official notice →- 2025-06-25
US
Dexcom, Inc. recalls Dexcom One+ Continuous Glucose Monitoring System, SKU(Part Number): STK-D7-013 (MT28285-1), STK-D7-014 (MT28285-1), STK-D7-103 (MT28285-2)
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low…
FDA · DeviceDexcom, Inc.worldwide distribution - US Nationwide and the countries of Andorra, Argentina…Official notice →- 2025-06-11
US
Dexcom, Inc. recalls Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system
A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are…
FDA · DeviceDexcom, Inc.Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of:…Official notice →- 2025-02-26
US
Dexcom, Inc. recalls Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OR-001
Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net…
FDA · DeviceDexcom, Inc.Worldwide - US Nationwide distribution in the states of OH, TX, PA, MN, SC,…Official notice →