Drugs & medicalClass I

Dexcom, Inc. recalls Dexcom One+ Continuous Glucose Monitoring System, SKU(Part Number): STK-D7-013 (MT28285-1), STK-D7-014 (MT28285-1), STK-D7-103 (MT28285-2)

Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low…

Read the official notice →
Date
2025-06-25
Country
United States
Brand
Dexcom
Product
Dexcom One+ Continuous Glucose Monitoring System, SKU(Part Number): STK-D7-013 (MT28285-1), STK-D7-014 (MT28285-1), STK-D7-103 (MT28285-2)
Classification
Class I
Units
worldwide distribution - US Nationwide and the countries of Andorra, Argentina…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Dexcom recalls

  1. 2026-07-01
  2. US
    Drugs & medicalClass II

    Dexcom, Inc. recalls Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog…

    Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC.…

    FDA · DeviceDexcom, Inc.Dexcom does not have distribution records for this product due to its…Official notice →
  3. 2026-05-20
  4. US
    Drugs & medicalClass II

    Dexcom, Inc. recalls Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software…

    A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV…

    FDA · DeviceDexcom, Inc.SW12300 & SW13355 Dexcom G7 iOS and watchOS app versions 2.9.0 through 2.11.2:…Official notice →
  5. 2025-11-05
  6. US
    Drugs & medicalClass I

    Dexcom, Inc. recalls Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number:…

    A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could…

    FDA · DeviceDexcom, Inc.This product was distributed throughout the US. G6 Android US v1.15.0 SW11678…Official notice →
  7. 2025-09-10
  8. US
    Drugs & medicalClass I

    Dexcom, Inc. recalls Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software…

    The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App…

    FDA · DeviceDexcom, Inc.SW12299 G7 Android CGM App Worldwide Distribution - US Nationwide and the…Official notice →
  9. 2025-06-11
  10. US
    Drugs & medicalClass II

    Dexcom, Inc. recalls Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system

    A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are…

    FDA · DeviceDexcom, Inc.Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of:…Official notice →
  11. 2025-02-26
  12. US
    Drugs & medicalClass II

    Dexcom, Inc. recalls Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OR-001

    Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net…

    FDA · DeviceDexcom, Inc.Worldwide - US Nationwide distribution in the states of OH, TX, PA, MN, SC,…Official notice →
All 7 Dexcom recalls →

Other drugs & medical recalls in the United States

  1. 2025-06-25
  2. US
    Drugs & medicalClass II

    Guangzhou Pluslife Biotech Co., Ltd. recalls Brand Name: Pluslife SARS-CoV-2 Card Product Name: SARS-CoV-2 Nucleic Acid Testing Card Model/Catalog Number: RM1010202…

    Devices were distributed in the U.S. market without obtaining the required FDA premarket authorization (e.g., 510(k) clearance or PMA).

    FDA · DeviceGuangzhou Pluslife Biotech Co., Ltd.Worldwide distribution: US (nationwide) and OUS (foreign) to countries of:…Official notice →
  3. US
    Drugs & medicalClass II

    Orthofix Srl recalls REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile

    Retrospective; Their is the potential that implantable intramedullary lengthening nails may bend at the distal tail during insertion resulting in having to be replaced.

    FDA · DeviceOrthofix SrlUS: Unknown, OUS: UnknownOfficial notice →
  4. US
    Drugs & medicalClass III

    HANGZHOU ACOSOUND TECHNOLOGY CO., LTD. recalls AcoSound. Model Number: LW12-BTE-M

    Three issues 1. Label error: The manufacturer information in the labeling is not correct, the manufacturer's name is mislabeled as "Developed by AcoSound," Correct should be "Manufactured by Hangzhou AcoSound." And the…

    FDA · DeviceHANGZHOU ACOSOUND TECHNOLOGY CO., LTD.US Nationwide distribution in the state of New York.Official notice →
  5. US
    Drugs & medicalClass II

    Spectral Instruments Inc recalls Ami HTX.

    The X-ray safety and manufacturing date labels affixed to the Lago X/ AMI HTX instrument do not correspond exactly to the labeling stipulated by the FDA / CDRH.

    FDA · DeviceSpectral Instruments IncUSOfficial notice →
  6. US
    Drugs & medicalClass I

    Cook Incorporated recalls Beacon Tip 5.0 Fr Angiographic catheters are designed to facilitate diagnostic and therapeutic procedures. Catheters are available in…

    Cook Medical identified that catheters supplied in the affected device lots may experience tip separation.

    FDA · DeviceCook IncorporatedWorldwide distribution.Official notice →
  7. US
    Drugs & medicalClass I

    Umary-USA.com recalls UNAVY ACIDO HIALURONICO, Conenido: 30 capletas de 850mg, distribuido por: EREN NATURAL S.A. DE C.V, Zappan, Jal. C.P, UPC 7 503046 054295

    Marketed without approved NDA/ANDA- Laboratory analysis found product to be tainted with undeclared ingredients: Dexamethasone, Diclofenac and Omeprazole.

    FDA · DrugUmary-USA.comU.S. Nationwide via internet.Official notice →