Drugs & medicalClass II

Beckman Coulter Mishima K.K. recalls Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 -…

Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of…

Read the official notice →
Date
2025-06-11
Country
United States
Brand
Beckman Coulter Mishima K.K
Product
Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 - 179433670
Classification
Class II
Units
Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Beckman Coulter Mishima K.K recalls

  1. 2026-10-07
  2. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965…

    Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA,…Official notice →
  3. 2026-03-18
  4. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281

    A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases…

    FDA · DeviceBeckman Coulter Mishima K.K.US Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC,…Official notice →
  5. 2025-06-11
  6. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 -…

    A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing.…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX,…Official notice →
  7. 2024-08-28
  8. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in…

    Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL,…Official notice →
  9. 2024-08-07
  10. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in…

    There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL,…Official notice →
All 6 Beckman Coulter Mishima K.K recalls →

Other drugs & medical recalls in the United States

  1. 2025-06-11
  2. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running software version 1.20

    A new software update (v1.20) is introducing error codes for the Dxl 9000 which prevents the assay from being performed and a delay in results from being generated, and may require field service to resolve the delay.

    FDA · DeviceBeckman Coulter, Inc.Worldwide Distribution: US (Nationwide) and OUS (Global) to countries of:…Official notice →
  3. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V

    Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed,…

    FDA · DeviceOlympus Corporation of the AmericasWorldwide Distribution: US (Nationwide) and OUS (International) countries to:…Official notice →
  4. US
    Drugs & medicalClass II

    Healthmark Industries Co., Inc. recalls ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample)

    During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries…

    FDA · DeviceHealthmark Industries Co., Inc.US Nationwide and Internationally to countries of: Canada, Malaysia.Official notice →
  5. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 -…

    A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing.…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX,…Official notice →
  6. US
    Drugs & medicalClass II

    Dexcom, Inc. recalls Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system

    A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are…

    FDA · DeviceDexcom, Inc.Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of:…Official notice →
  7. US
    Drugs & medicalClass II

    VIRTUAL INCISION CORPORATION recalls MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version…

    The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the…

    FDA · DeviceVIRTUAL INCISION CORPORATIONUS Distribution to States: CA, FL, NEOfficial notice →
See all drugs & medical recalls →