Beckman Coulter Mishima K.K. recalls DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in…

Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will…

Read the official notice →
Date
2024-08-28
Country
United States
Brand
Beckman Coulter Mishima K.K
Product
DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination…
Classification
Class II
Units
Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Beckman Coulter Mishima K.K recalls

  1. 2026-10-07
  2. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965…

    Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA,…Official notice →
  3. 2026-03-18
  4. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281

    A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases…

    FDA · DeviceBeckman Coulter Mishima K.K.US Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC,…Official notice →
  5. 2025-06-11
  6. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 -…

    Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX,…Official notice →
  7. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 -…

    A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing.…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX,…Official notice →
  8. 2024-08-07
  9. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in…

    There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL,…Official notice →
All 6 Beckman Coulter Mishima K.K recalls →

Other drugs & medical recalls in the United States

  1. 2024-08-28
  2. US
    Drugs & medicalClass II

    NeuroLogica Corporation recalls GM60A. Digital Diagnostic Mobile X-Ray System.

    Potential for the anti-fall system of the device arm to fail.

    FDA · DeviceNeuroLogica CorporationUS Nationwide distribution in the states of AR, CT, GA, IN, KS, MA, MD, MO,…Official notice →
  3. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls LINER, SOFT, SUCTION CANISTER, 1500CC, REF DYNDSCL1500

    A slight dimensional variation which has the potential for increased difficulty fitting into outer canister.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide - US Nationwide and the countries of Panama, Mexico, Canada, Qatar,…Official notice →
  4. US
    Drugs & medicalClass II

    Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235- Designed to produce and deliver a proton beam for the treatment of patients with localized…

    Temporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues

    FDA · DeviceIon Beam Applications S.A.IL,LA, NJ, MI,OK, TN, TX, VA, WA Foreign: Czech Republic France Germany Italy…Official notice →
  5. US
    Drugs & medicalClass II

    Stryker Sustainability Solutions recalls Disposable 5 Lead Cable and Lead Wire System, 50 inches, REF: LW-309DS50/5A

    Disposable 5 lead cable and lead wire system package that should contain lead systems, lead telemetry systems and a dual connect cable, may in fact contain, different ECG Leads, which are not approved for reprocessing,…

    FDA · DeviceStryker Sustainability SolutionsUS Nationwide distribution in the states of FL, WY, AZ, CA, LA, TX, MI, OR,…Official notice →
  6. US
    Drugs & medicalClass II

    Intrinsic Therapeutics, Inc. recalls Barricaid Annular Closure Impactor, REF: BAR-IMPACTOR, P/N: 400906-A

    At least two units have been found to be missing a weld.

    FDA · DeviceIntrinsic Therapeutics, Inc.US Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Preat Corp recalls Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF 9000703 Straumann…

    Due to a manufacturing issue, the product has a out of specification rotation of the digital analog used to visually inspect dental abutments and restorations.

    FDA · DevicePreat CorpU.S. Nationwide distribution in the states of AK, AZ, FL, GA, IL, LA, MD, MI,…Official notice →
See all drugs & medical recalls →