Drugs & medicalClass II

Preat Corp recalls Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF 9000703 Straumann…

Due to a manufacturing issue, the product has a out of specification rotation of the digital analog used to visually inspect dental abutments and restorations.

Read the official notice →
Date
2024-08-28
Country
United States
Brand
Preat
Product
Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF 9000703 Straumann…
Classification
Class II
Units
U.S. Nationwide distribution in the states of AK, AZ, FL, GA, IL, LA, MD, MI,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Preat recalls

  1. 2025-09-10
  2. US
    Drugs & medicalClass II

    Preat Corp recalls NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662

    Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the…

    FDA · DevicePreat CorpUS Nationwide distribution in the states of NY, FL, PA, CA, OH.Official notice →
  3. 2025-07-16
  4. US
    Drugs & medicalClass II

    Preat Corp recalls Preat, Astra EV-compatible 4.2mm Digital Analog REF: 9006603 (single) and 9006603-10(10 pack), Rx Only

    Due to manufacturing error, digital analog rotational feature is 180 degrees off.

    FDA · DevicePreat CorpUS Nationwide distribution in the states of AL, CA, FL, GA, KY, MD, MN, NY,…Official notice →
  5. 2025-06-25
  6. US
    Drugs & medicalClass II

    Preat Corp recalls Biomet 3i Certain-compatible 4.1 mm Milled Ti Abutment REF 9001767-F Preat milled Abutments are intended to be used in conjunction with…

    Due to manufacturing issues, abutment products were manufactured with screw seat located too high, which may cause the mating screw to have a shorter engagement length into the associated implant.

    FDA · DevicePreat CorpU.S. Nationwide distribution in the states of AZ, CA, GA, MO, NJ, NY, OH, PA,…Official notice →
  7. 2024-07-17
  8. US
    Drugs & medicalClass II

    Preat Corp recalls 9006767 Astra EV 4.8mm Milled Titanium Abutment Milled abutments are too be used in conjunction with an endosseous dental implant fixture…

    Due to incorrect milled titanium abutment being packaged and labeled.

    FDA · DevicePreat CorpU.S.: CA, MO, and NV. O.U.S.: NoneOfficial notice →
  9. 2024-03-27
  10. US
    Drugs & medicalClass II

    Preat Corp recalls Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967

    Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant.

    FDA · DevicePreat CorpU.S. Nationwide distribution in the states of AZ, OH and PA.Official notice →
  11. 2023-11-01
  12. US
    Drugs & medicalClass II

    Preat Corp recalls Preat REF 9007121, Neodent GM x 2mm Healing Abutment, Rx Only, Non-Sterile, Single Use Only. Dental

    The healing abutments contain a dimensional condition that may lead to a gap between the healing abutment and the implant allowing for microleakage

    FDA · DevicePreat CorpUS: FL, CA, OR, TX, OH, CA, SC, OR, NJ, SC, OUS: NoneOfficial notice →
All 9 Preat recalls →

Other drugs & medical recalls in the United States

  1. 2024-08-28
  2. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in…

    Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL,…Official notice →
  3. US
    Drugs & medicalClass II

    Stryker Sustainability Solutions recalls Disposable 5 Lead Cable and Lead Wire System, 50 inches, REF: LW-309DS50/5A

    Disposable 5 lead cable and lead wire system package that should contain lead systems, lead telemetry systems and a dual connect cable, may in fact contain, different ECG Leads, which are not approved for reprocessing,…

    FDA · DeviceStryker Sustainability SolutionsUS Nationwide distribution in the states of FL, WY, AZ, CA, LA, TX, MI, OR,…Official notice →
  4. US
    Drugs & medicalClass II

    Intrinsic Therapeutics, Inc. recalls Barricaid Annular Closure Impactor, REF: BAR-IMPACTOR, P/N: 400906-A

    At least two units have been found to be missing a weld.

    FDA · DeviceIntrinsic Therapeutics, Inc.US Nationwide distribution.Official notice →
  5. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD BACTEC MGIT 960 PZA Kit)- IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to…

    May intermittently produce falsely resistant results for pyrazinamide (PZA) during susceptibility testing of Mycobacterium tuberculosis isolates.

    FDA · DeviceBecton Dickinson & Co.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  6. US
    Drugs & medicalClass II

    ROi CPS LLC recalls Regard Kit containing the Microtek decanter

    Decanters may have pin holes and tears in sterile packaging that cannot be identified visually.

    FDA · DeviceROi CPS LLCUS Nationwide distribution in the state of MO.Official notice →
  7. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and a total protein test…

    The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT…

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Nationwide. Foreign: Australia Belgium Bermuda Brazil Canada L3R 0B8 Chile…Official notice →
See all drugs & medical recalls →