Drugs & medicalClass II

Ortho-Clinical Diagnostics, Inc. recalls The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and a total protein test…

The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT…

Read the official notice →
Date
2024-08-28
Country
United States
Brand
Ortho-Clinical Diagnostics
Product
The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and a total protein test side by side…
Classification
Class II
Units
Nationwide. Foreign: Australia Belgium Bermuda Brazil Canada L3R 0B8 Chile…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Ortho-Clinical Diagnostics recalls

  1. 2025-01-29
  2. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, INc. recalls Brand Name: VITROS Product Name: VITROS XT 7600 Integrated System Model/Catalog Number: 6844461 (New), 6272222 (Certified),…

    The VITROS XT 7600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired…

    FDA · DeviceOrtho-Clinical Diagnostics, INc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  3. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, INc. recalls Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: 6802413 (New), 6802415…

    The VITROS 5600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired…

    FDA · DeviceOrtho-Clinical Diagnostics, INc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  4. 2024-12-04
  5. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in…

    May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to…

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of AZ, FL, ID, IL, KS,…Official notice →
  6. 2024-08-21
  7. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2

    It was confirmed that fibrinogen in patient plasma samples precipitates out of solution upon the addition of folate stabilizer reagent as part of the pre-treatment process causing the increase in "TM5-4MB" condition…

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  8. 2024-07-10
  9. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products FS Diluent Pack 3, Model/Catalog Number: 6801754. Product Description: VITROS Chemistry Products FS…

    VITROS Chemistry Products FS Diluent Pack 3, Lot 01-1645 may cause results to be negatively biased when used to dilute patient samples tested with VITROS Chemistry Products CRP Slides.

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of FL, IL, NY, PA and the…Official notice →
  10. US
    Drugs & medicalClass III

    Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS…

    Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration date for Lot 2953 is 24-Jul-2024. If used past its true expiration and the customer does not have their Assay Data Disk…

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide including Puerto Rico and the countries…Official notice →
All 44 Ortho-Clinical Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2024-08-28
  2. US
    Drugs & medicalClass II

    Preat Corp recalls Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF 9000703 Straumann…

    Due to a manufacturing issue, the product has a out of specification rotation of the digital analog used to visually inspect dental abutments and restorations.

    FDA · DevicePreat CorpU.S. Nationwide distribution in the states of AK, AZ, FL, GA, IL, LA, MD, MI,…Official notice →
  3. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD BACTEC MGIT 960 PZA Kit)- IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to…

    May intermittently produce falsely resistant results for pyrazinamide (PZA) during susceptibility testing of Mycobacterium tuberculosis isolates.

    FDA · DeviceBecton Dickinson & Co.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  4. US
    Drugs & medicalClass II

    ROi CPS LLC recalls Regard Kit containing the Microtek decanter

    Decanters may have pin holes and tears in sterile packaging that cannot be identified visually.

    FDA · DeviceROi CPS LLCUS Nationwide distribution in the state of MO.Official notice →
  5. US
    Drugs & medicalClass II

    BioFire Diagnostics, LLC recalls BIOFIRE FILMARRAY Pneumonia Panel plus, REF: RFIT-ASY-0143

    Pneumonia panels have an increased risk of control failures due to an issue in the manufacturing process, which could lead to delayed results for the customer.

    FDA · DeviceBioFire Diagnostics, LLCWorldwide - US Nationwide distribution in the states of WV, IN, AL and the…Official notice →
  6. US
    Drugs & medicalClass II

    Abbott Laboratories recalls Alinity hq Analyzer, REF: 09P68-01

    1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating…

    FDA · DeviceAbbott LaboratoriesWorldwide - US Nationwide distribution in the states of NY, OK, NC and the…Official notice →
  7. US
    Drugs & medicalClass II

    Alcon Research LLC recalls Product Desc: Product Name: CLEARCUT SB INTREPID 2.2 ANG Model/Catalog Number: 8065992561 Software Version: N/A Product Description:…

    Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives.

    FDA · DeviceAlcon Research LLCDomestic: Nationwide Distribution Countries Affected: Argentina, Australia,…Official notice →