Ortho-Clinical Diagnostics recalls
44 official recalls and safety alerts involving Ortho-Clinical Diagnostics since 2021-10, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Ortho-Clinical Diagnostics or the relevant agency.
- 2025-09-24
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products LAC Slides. Catalog Number: 8150112. 1 Sales Unit = contains 5 cartridges of 18 slides, for a total…
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J…
- 2025-08-27
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261;
Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2025-07-23
US
Ortho-Clinical Diagnostics, Inc. recalls Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806…
Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution;…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD,…Official notice →- 2025-07-09
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package…
Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Ortho-Clinical Diagnostics, Inc. recalls Brand Name: VITROS; Product Name: VITROS XT 7600 Integrated System; Model/Catalog Number: (1) 6844461, (2) 6272222; Software…
a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide Distribution: US (Nationwide) and OUS (International) to countries…Official notice →US
Ortho-Clinical Diagnostics, Inc. recalls Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915
A software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide Distribution: US (Nationwide) and OUS (International) to countries…Official notice →- 2025-06-18
US
Ortho-Clinical Diagnostics, Inc. recalls Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software Version:…
It was determined that there could be a false positive rate of up to 16.7% for the affected lots associated with the Bordetella parapertussis (BPP) results (only) of the assay.
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Please see the attached list "Consignee List_CL2025-120." Note: This file…Official notice →- 2025-06-04
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, of 18 slides per cartridge, for a total of 90 slides.
Product was incorrectly assigned GEN 88 instead of GEN 83 on product carton, foil wrapper, and cartridge (cart) label.
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of AR, AZ, CO, FL, GA,…Official notice →- 2025-01-29
US
Ortho-Clinical Diagnostics, INc. recalls Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System Model/Catalog Number: 6802783 (New), 6802914…
The VITROS 3600 Immunodiagnostic System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired…
FDA · DeviceOrtho-Clinical Diagnostics, INc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Ortho-Clinical Diagnostics, INc. recalls Brand Name: VITROS Product Name: VITROS XT 7600 Integrated System Model/Catalog Number: 6844461 (New), 6272222 (Certified),…
The VITROS XT 7600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired…
FDA · DeviceOrtho-Clinical Diagnostics, INc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Ortho-Clinical Diagnostics, INc. recalls Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: 6802413 (New), 6802415…
The VITROS 5600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired…
FDA · DeviceOrtho-Clinical Diagnostics, INc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2024-12-04
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in…
May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of AZ, FL, ID, IL, KS,…Official notice →- 2024-08-28
US
Ortho-Clinical Diagnostics, Inc. recalls The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and a total protein test…
The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Nationwide. Foreign: Australia Belgium Bermuda Brazil Canada L3R 0B8 Chile…Official notice →- 2024-08-21
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2
It was confirmed that fibrinogen in patient plasma samples precipitates out of solution upon the addition of folate stabilizer reagent as part of the pre-treatment process causing the increase in "TM5-4MB" condition…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2024-07-10
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products FS Diluent Pack 3, Model/Catalog Number: 6801754. Product Description: VITROS Chemistry Products FS…
VITROS Chemistry Products FS Diluent Pack 3, Lot 01-1645 may cause results to be negatively biased when used to dilute patient samples tested with VITROS Chemistry Products CRP Slides.
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of FL, IL, NY, PA and the…Official notice →US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS…
Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration date for Lot 2953 is 24-Jul-2024. If used past its true expiration and the customer does not have their Assay Data Disk…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide including Puerto Rico and the countries…Official notice →- 2024-06-19
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Anti HBs Calibrators. The VITROS Anti-HBs Calibrator Pack contains 3 X labelled Calibrator…
Potential for positively biased results due to signal reduction over shelf life.
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →- 2024-06-12
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Vitamin B12 Range Verifier - In vitro diagnostic use ly. For use in verifying the calibration…
The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 1200 RV Level 2 may cause range verification failure. The issue is detectable by the user when the product fails to meet the…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide and the country of Columbia.Official notice →- 2024-04-03
US
Ortho-Clinical Diagnostics recalls VITROS Performance Verifier I
The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing…
FDA · DeviceOrtho-Clinical DiagnosticsWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2024-03-13
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (refurbished)
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA,…Official notice →US
Ortho-Clinical Diagnostics, Inc. recalls VITROS 5600 Integrated System (New and Refurbished)
Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VITROS¿ 3600 Immunodiagnostic Systems and VITROS¿ 5600/XT 7600 Integrated Systems which may lead to erroneous test results.
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2024-01-31
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Anti-HBs Quantitative Reagent Pack, Catalog No. 6801925 VITROS Immunodiagnostic Products…
Affected lots may experience increased calibration failures or an increase in falsely elevated results.
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Domestic distribution nationwide. OUS distribution to Bermuda, Australia,…Official notice →- 2024-01-17
US
Ortho-Clinical Diagnostics, Inc. recalls QMS Tacrolimus Calibrators, Product code 10015573
Ortho-Clinical Diagnostics, Inc. is recalling QMS Tacrolimus Calibrators, Lot 74859619 expiry date: May 11, 2024, due to improper storage that is expected to degrade the quality of the calibrator and render it…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Domestic distribution to IL, IN, FL, NJ, NM, TX. International distribution to…Official notice →- 2023-12-20
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS XT 3400 Chemistry System Product Code 6844458 Running Software Version 3.8.0
During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen…
- 2023-12-13
US
Ortho-Clinical Diagnostics, INc. recalls VITROS Chemistry Products ECO2 Slides-For in vitro diagnostic use only. VITROS Chemistry Products ECO2 Slides quantitatively…
Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coating 3598 of VITROS ECO2 slides included carts which may contain consecutive slides that may not function as intended.…
- 2023-11-15
US
Ortho-Clinical Diagnostics, INc. recalls VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18…
Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed…
FDA · DeviceOrtho-Clinical Diagnostics, INc.US States: CA, CO, FL, GA, IL, IN, KS, NC, NY, TN, TX, VT. Canada, France,…Official notice →- 2023-10-11
US
Ortho-Clinical Diagnostics, Inc. recalls Proboscis/Piston Assembly, Part Number J55375, for VITROS 5,1 FS Chemistry System (Product Code 6801375), VITROS 5,1 Chemistry…
Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may not perform as intended due to an…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution.Official notice →- 2023-09-27
US
Ortho-Clinical Diagnostics, INc. recalls VITROS XT Chemistry Products ALKP Slides
Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in…
FDA · DeviceOrtho-Clinical Diagnostics, INc.Worldwide Distribution: US (nationwide) and OUS (foreign) distribution to…Official notice →- 2023-08-16
US
ORTHO-CLINICAL DIAGNOSTICS recalls VITROS Immunodiagnostic Products Troponin I ES Reagent Pack
Reagent packs could potentially contain incorrect wells, which will generate incorrect, and potentially false negative, results.
FDA · DeviceORTHO-CLINICAL DIAGNOSTICSWorldwide distribution - US Nationwide and the country of Canada.Official notice →- 2023-08-02
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products NT-proBNP II Range Verifiers-For in vitro diagnostic use only. For use in verifying the…
May not meet the stability specifications for shelf life outlined in the Instructions For Use (IFU).
FDA · DeviceOrtho-Clinical Diagnostics, Inc.US Nationwide distribution in the state of CA.Official notice →- 2023-07-19
US
Ortho-Clinical Diagnostics recalls VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- Used to…
Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected…
FDA · DeviceOrtho-Clinical DiagnosticsWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2023-06-07
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Calibrator Kit 11 VANC Reagent Generation (GEN) 50 with affected Assay Data Disk (ADD) for Data…
Unavailable to calibrate VITROS Chemistry Products VANC Reagent, Generation (GEN) 50, results in delay of testing
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Nationwide Foreign: Brazil, Chile, ColombiaOfficial notice →- 2023-05-31
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products HbA1c Reagent Kit- IVD for the quantitative determination of percent glycated hemoglobin A1c…
Potential for both unexpected positive and negative HbA1c results, may result in misdiagnosis
FDA · DeviceOrtho-Clinical Diagnostics, Inc.US Nationwide- Worldwide Distribution. Foreign: Australia Belgium Brazil…Official notice →- 2023-05-10
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS 5600 Integrated System- For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a…
MicroTip Pack Opener Assembly Potentially Not Removing or Replacing MicroTip Pack Caps on VITROS¿ 5600 and XT 7600 Integrated Systems, results may be delayed results if a pack becomes unusable
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Nationwide including Puerto Rico and Bermuda Foreign: Australia Belgium Brazil…Official notice →- 2023-04-26
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Anti-SARSCoV- 2 Total N Antibody Calibrators programmed onto the Assay Data Disks (ADD) Data…
Incorrect Number of Calibrator Levels (2) Set for the VITROS Immunodiagnostics Products Anti-SARS-CoV-2 Total N Antibody Calibrators in Assay Data Disks (ADD), resulting in delayed patients results
FDA · DeviceOrtho-Clinical Diagnostics, Inc.US Nationwide distribution in the states of AL, AZ, CA, GA, NY.Official notice →- 2023-01-11
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Intact PTH Reagent Pack- IVD For the quantitative measurement of intact parathyroid hormone…
Lower than expected iPTH patient and Quality Control (QC) results, reported negatively biased iPTH results- an average -12% shift in patient sample results
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2022-11-16
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use only. Product Code:…
Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide Distribution: US (Nationwide) and Countries (Foreign) including:…Official notice →- 2022-09-14
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461
Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer.…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2022-07-20
US
Ortho-Clinical Diagnostics, Inc. recalls Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS¿ ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS¿…
On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU).
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Colombia, India,…Official notice →- 2022-05-25
US
Ortho-Clinical Diagnostics, Inc recalls VITROS Immunodiagnostic Products LH Reagent Pack Product Code: 1350198
Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at…
FDA · DeviceOrtho-Clinical Diagnostics, IncWorldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda…Official notice →- 2022-03-16
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.
The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Domestic distribution nationwide. Foreign distribution to Canada.Official notice →- 2022-01-26
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products High Sample Diluent A Reagent Pack
Lower than expected results for VITROS¿ Immunodiagnostic Products TSH Reagent Pack when diluted with VITROS¿ Immunodiagnostic Products High Sample Diluent A Reagent Pack Lot 2190.
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide distribution in the states of AL, AR,…Official notice →- 2021-11-24
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Anti-HBc IgM Controls, Product Code 6801427 For use in monitoring the performance of the VITROS…
Ortho Clinical Diagnostics confirmed a complaint that some VITROS Immunodiagnostic Products Anti-HBc IgM Controls Lot 1690 Controls Values booklet contained information for the incorrect product. While the accuracy of…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.US (Domestic) distribution to states of: AZ, CA, CO, CT, FL, GA, LA, MA, MO,…Official notice →- 2021-10-27
US
Ortho Clinical Diagnostics Inc recalls VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Quantitative Controls (IVD)- For use in monitoring the performance of the…
VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Quantitative Controls Outside of Expected Control Range, potential for delayed results
FDA · DeviceOrtho Clinical Diagnostics IncWorldwide Distribution: US (Nationwide) in states of: AL, AR, AZ, CA, CO, DC,…Official notice →