Ortho-Clinical Diagnostics, Inc. recalls VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461
Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer.…
Read the official notice →- Date
- 2022-09-14
- Country
- United States
- Brand
- Ortho-Clinical Diagnostics
- Product
- VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Australia,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Ortho-Clinical Diagnostics recalls
- 2023-04-26
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Anti-SARSCoV- 2 Total N Antibody Calibrators programmed onto the Assay Data Disks (ADD) Data…
Incorrect Number of Calibrator Levels (2) Set for the VITROS Immunodiagnostics Products Anti-SARS-CoV-2 Total N Antibody Calibrators in Assay Data Disks (ADD), resulting in delayed patients results
FDA · DeviceOrtho-Clinical Diagnostics, Inc.US Nationwide distribution in the states of AL, AZ, CA, GA, NY.Official notice →- 2023-01-11
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Intact PTH Reagent Pack- IVD For the quantitative measurement of intact parathyroid hormone…
Lower than expected iPTH patient and Quality Control (QC) results, reported negatively biased iPTH results- an average -12% shift in patient sample results
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2022-11-16
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use only. Product Code:…
Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide Distribution: US (Nationwide) and Countries (Foreign) including:…Official notice →- 2022-07-20
US
Ortho-Clinical Diagnostics, Inc. recalls Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS¿ ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS¿…
On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU).
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Colombia, India,…Official notice →- 2022-05-25
US
Ortho-Clinical Diagnostics, Inc recalls VITROS Immunodiagnostic Products LH Reagent Pack Product Code: 1350198
Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at…
FDA · DeviceOrtho-Clinical Diagnostics, IncWorldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda…Official notice →- 2022-03-16
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.
The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Domestic distribution nationwide. Foreign distribution to Canada.Official notice →
Other drugs & medical recalls in the United States
- 2022-09-14
US
Abbott Molecular, Inc. recalls Abbott Alinity m SARS-CoV-2 CTRL Kit, REF 09N78-085, each kit contains 12 x 1.3mL CTRL - and 12 x 1.3 mL CTRL +, For In Vitro Diagnostic…
Reports received of an increase of message code 9186 (internal control failed) when using the kit.
FDA · DeviceAbbott Molecular, Inc.Distribution was made to AZ, CA, CT, FL, GA, IL, KS, KY, LA, MA, MI, MN, MO,…Official notice →US
OrthoPediatrics Corp recalls Orthex External Fixation Wire Sharp Bayonet 1.8 x 400 mm Model: K400-18BN
Manufacturing variance near transition of tip is increasing the likelihood the wire tip may fracture. The fractured tip may be retained in the patient. Retrieval attempts may be made which could result in a delay of…
FDA · DeviceOrthoPediatrics CorpUS Nationwide Distribution: AZ, AL, CA, FL, FL, MD, MO, NC, NY, TX Foreign:…Official notice →US
Biosense Webster, Inc. recalls CARTO VIZIGO 8.5F Bi-Directional Guiding Sheath, REF: D138501. Used to introduce various cardiovascular catheters into the heart.
Outside packaging for a device indicated for introducing cardiovascular catheters into the heart, sustained water damage, which may have affected the sterile barrier of the product inside. Devices with a compromised…
FDA · DeviceBiosense Webster, Inc.US Nationwide distribution in the states of AL, GA, AZ, TX, MA, VA, NY, ME,…Official notice →US
US
Philips Respironics, Inc. recalls Philips BiPAP A30, Rx Only. To provide noninvasive ventilation support for patients with obstructive sleep apnea syndrome (OSA) and…
Motor assemblies manufactured with nonconforming plastic that contributes to off-gassing of VOCs.
FDA · DevicePhilips Respironics, Inc.Domestic: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA,…Official notice →US
CIPLA recalls Difluprednate Ophthalmic Emulsion, 0.05%, 5 mL bottle, RX only, Manufactured for Exelan Pharmecuticals, Inc., Boca Raton, FL 33432, NDC…
Lack of Assurance of Sterility: customer complaint for defective container where breakage of the protective cap exposes tip of eye drop which could compromise sterility.