Drugs & medicalClass II

Abbott Molecular, Inc. recalls Abbott Alinity m SARS-CoV-2 CTRL Kit, REF 09N78-085, each kit contains 12 x 1.3mL CTRL - and 12 x 1.3 mL CTRL +, For In Vitro Diagnostic…

Reports received of an increase of message code 9186 (internal control failed) when using the kit.

Read the official notice →
Date
2022-09-14
Country
United States
Brand
Abbott Molecular
Product
Abbott Alinity m SARS-CoV-2 CTRL Kit, REF 09N78-085, each kit contains 12 x 1.3mL CTRL - and 12 x 1.3 mL CTRL +, For In Vitro Diagnostic…
Classification
Class II
Units
Distribution was made to AZ, CA, CT, FL, GA, IL, KS, KY, LA, MA, MI, MN, MO,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abbott Molecular recalls

  1. 2023-02-01
  2. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing,…

    The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration…

    FDA · DeviceAbbott Molecular, Inc.xOfficial notice →
  3. 2022-12-28
  4. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Abbott Alinity m Resp-4-Plex Amp Kit, Part Number 09N79-096

    Reports of an increase in reactive negative controls and false positive results with certain lots of Alinity M Resp-4-Plex AMP kits.

    FDA · DeviceAbbott Molecular, Inc.WorldwideOfficial notice →
  5. 2022-11-30
  6. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use

    Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flapper was found to not open when the Systems Solution drawer was closed and…

    FDA · DeviceAbbott Molecular, Inc.Worldwide distribution.Official notice →
  7. 2022-08-24
  8. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls The Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use.

    Incorrect optical calibration

    FDA · DeviceAbbott Molecular, Inc.There was only foreign distribution to Italy.Official notice →
  9. 2022-04-13
  10. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m HBV AMP Kit (US and CE)

    There is a potential for misquantitation high results for negative samples.

    FDA · DeviceAbbott Molecular, Inc.US consignees - AZ, CA, IL, IN, MA, MI, MO, NH, NJ, NY, OH, SD, TX, and VA OUS…Official notice →
  11. US
    Drugs & medicalClass II

    Abbott Molecular, Inc. recalls Alinity m HBV Application Specification File (US and CE)

    There is a potential for misquantitation high results of negative samples.

    FDA · DeviceAbbott Molecular, Inc.US consignees - AZ, CA, IL, IN, MA, MI, MO, NH, NJ, NY, OH, SD, TX, and VA OUS…Official notice →
All 21 Abbott Molecular recalls →

Other drugs & medical recalls in the United States

  1. 2022-09-14
  2. US
    Drugs & medicalClass II

    Medline Industries Inc. recalls Medline CHEST TUBE REORDER NO: DYNJ36762A

    Due to a non-sterile component (tape) being place within the sterile portion of the Chest Tube Convenience Kit

    FDA · DeviceMedline Industries Inc.U.S. Nationwide distribution in the state of AL.Official notice →
  3. US
    Drugs & medicalClass II

    ICU Medical, Inc. recalls Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIVE PORT, DRY SPIKE/CL-12; APPX…

    Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic…

    FDA · DeviceICU Medical, Inc.Worldwide distribution - US Nationwide distribution including in the states of…Official notice →
  4. US
    Drugs & medicalClass II

    BALT USA, LLC recalls Optima Coil System Model OPTI0520CSF10. For endovascular embolization.

    Products were released to customers prior to the approval and release of a test report authorizing the use of an existing material from an alternate supplier.

    FDA · DeviceBALT USA, LLCU.S.: AZ and FL O.U.S.: NoneOfficial notice →
  5. US
    Drugs & medicalClass II

    OrthoPediatrics Corp recalls Orthex External Fixation Wire Sharp Bayonet 1.8 x 400 mm Model: K400-18BN

    Manufacturing variance near transition of tip is increasing the likelihood the wire tip may fracture. The fractured tip may be retained in the patient. Retrieval attempts may be made which could result in a delay of…

    FDA · DeviceOrthoPediatrics CorpUS Nationwide Distribution: AZ, AL, CA, FL, FL, MD, MO, NC, NY, TX Foreign:…Official notice →
  6. US
    Drugs & medicalClass II

    Biosense Webster, Inc. recalls CARTO VIZIGO 8.5F Bi-Directional Guiding Sheath, REF: D138501. Used to introduce various cardiovascular catheters into the heart.

    Outside packaging for a device indicated for introducing cardiovascular catheters into the heart, sustained water damage, which may have affected the sterile barrier of the product inside. Devices with a compromised…

    FDA · DeviceBiosense Webster, Inc.US Nationwide distribution in the states of AL, GA, AZ, TX, MA, VA, NY, ME,…Official notice →
  7. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461

    Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer.…

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
See all drugs & medical recalls →