Ortho-Clinical Diagnostics, Inc. recalls VITROS 5600 Integrated System- For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a…
MicroTip Pack Opener Assembly Potentially Not Removing or Replacing MicroTip Pack Caps on VITROS¿ 5600 and XT 7600 Integrated Systems, results may be delayed results if a pack becomes unusable
Read the official notice →- Date
- 2023-05-10
- Country
- United States
- Brand
- Ortho-Clinical Diagnostics
- Product
- VITROS 5600 Integrated System- For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of…
- Classification
- Class II
- Units
- Nationwide including Puerto Rico and Bermuda Foreign: Australia Belgium Brazil…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Ortho-Clinical Diagnostics recalls
- 2023-07-19
US
Ortho-Clinical Diagnostics recalls VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- Used to…
Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected…
FDA · DeviceOrtho-Clinical DiagnosticsWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2023-06-07
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Calibrator Kit 11 VANC Reagent Generation (GEN) 50 with affected Assay Data Disk (ADD) for Data…
Unavailable to calibrate VITROS Chemistry Products VANC Reagent, Generation (GEN) 50, results in delay of testing
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Nationwide Foreign: Brazil, Chile, ColombiaOfficial notice →- 2023-05-31
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products HbA1c Reagent Kit- IVD for the quantitative determination of percent glycated hemoglobin A1c…
Potential for both unexpected positive and negative HbA1c results, may result in misdiagnosis
FDA · DeviceOrtho-Clinical Diagnostics, Inc.US Nationwide- Worldwide Distribution. Foreign: Australia Belgium Brazil…Official notice →- 2023-04-26
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Anti-SARSCoV- 2 Total N Antibody Calibrators programmed onto the Assay Data Disks (ADD) Data…
Incorrect Number of Calibrator Levels (2) Set for the VITROS Immunodiagnostics Products Anti-SARS-CoV-2 Total N Antibody Calibrators in Assay Data Disks (ADD), resulting in delayed patients results
FDA · DeviceOrtho-Clinical Diagnostics, Inc.US Nationwide distribution in the states of AL, AZ, CA, GA, NY.Official notice →- 2023-01-11
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Intact PTH Reagent Pack- IVD For the quantitative measurement of intact parathyroid hormone…
Lower than expected iPTH patient and Quality Control (QC) results, reported negatively biased iPTH results- an average -12% shift in patient sample results
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2022-11-16
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use only. Product Code:…
Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide Distribution: US (Nationwide) and Countries (Foreign) including:…Official notice →
Other drugs & medical recalls in the United States
- 2023-05-17
US
Camber Pharmaceuticals Inc. recalls Atovaquone Oral Suspension USP, 750 mg/5 mL, Packaged in 210 mL bottle, Manufactured for: Camber Pharmaceuticals, Inc., Piscataway,…
Microbial Contamination of Non-Sterile Product: Objectionable organism, identified as Bacillus cereus, found in product during testing of repackaged product.
- 2023-05-10
US
Inpeco S.A. recalls Storage and Retrieval Module (SRM)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP),…
Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either…
FDA · DeviceInpeco S.A.Worldwide distribution - US Nationwide and the countries of ARGENTINA,…Official notice →US
DeVilbiss Healthcare LLC recalls DEVILBISS HEALTHCARE 1025 SERIES OXYGEN CONCENTRATOR, Model Numbers 1025DS, 1025UK, 1025KS, and 1025KS-AR
The instruction manual is being replaced with an updated version that is in alignment with labeling requirements specified in the IEC-60101-1 standard.
US
Linkbio Corp. recalls LINK SLED Knee System: ALL POLY TIBIAL COMPONENT 7MM X 58MM Item Number: 15-2028/13
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal…
FDA · DeviceLinkbio Corp.NY, NV Foreign: Bulgaria, France, Italy, Norway, Poland, Sweden, Switzerland,…Official notice →US
DermLite LLC recalls DERMLITE DL4W, REF: DL4W, Rx Only MD, CE
Incorrect labeling; Package labeling contains a different serial number then the serial number on the product labeling.
FDA · DeviceDermLite LLCWorldwide - US Nationwide distribution in the states of CA, AZ, TX, FL, NE,…Official notice →US
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Cortisol 50T (Material Number 10995538), 250T (Material Number 10995537), and REF 250T (Material Number…
There is a negative bias with urine patient samples and urine cortisol Quality Control (QC) results intermittently out of range low.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide.Official notice →