Drugs & medicalClass II

Linkbio Corp. recalls LINK SLED Knee System: ALL POLY TIBIAL COMPONENT 7MM X 58MM Item Number: 15-2028/13

Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal…

Read the official notice →
Date
2023-05-10
Country
United States
Brand
Linkbio
Product
LINK SLED Knee System: ALL POLY TIBIAL COMPONENT 7MM X 58MM Item Number: 15-2028/13
Classification
Class II
Units
NY, NV Foreign: Bulgaria, France, Italy, Norway, Poland, Sweden, Switzerland,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Linkbio recalls

  1. 2026-04-22
  2. US
    Drugs & medicalClass II

    Linkbio Corp. recalls LinkBio CORE Workstation, a component of the CORE Shoulder System. Product Code: 87-9136.

    The CORE Workstation may display an incorrect "Planning Date " when viewing cases on the "Start Case" menu screen.

    FDA · DeviceLinkbio Corp.US Nationwide distribution in the state of Alabama, Florida, and Kansas.Official notice →
  3. 2023-06-07
  4. US
    Drugs & medicalClass II

    Linkbio Corp. recalls LINK SymphoKnee System L-Shaped Femoral Augment, Medial-Right/Lateral-Left, Sz 5-6 Tilastan, cemented. Intended for primary and revision…

    Contains the incorrect implant in the packaging. The packaging incorrectly contains the opposite side of the implant (Lateral-Right/Medial-Left instead of Medial-Right/Lateral-Left).

    FDA · DeviceLinkbio Corp.US Nationwide Distribution - AL, GA, IL, IN, LA, NV, TXOfficial notice →
All 3 Linkbio recalls →

Other drugs & medical recalls in the United States

  1. 2023-05-10
  2. US
    Drugs & medicalClass II

    Inpeco S.A. recalls Storage and Retrieval Module (SRM)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP),…

    Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either…

    FDA · DeviceInpeco S.A.Worldwide distribution - US Nationwide and the countries of ARGENTINA,…Official notice →
  3. US
    Drugs & medicalClass II

    DeVilbiss Healthcare LLC recalls DEVILBISS HEALTHCARE 1025 SERIES OXYGEN CONCENTRATOR, Model Numbers 1025DS, 1025UK, 1025KS, and 1025KS-AR

    The instruction manual is being replaced with an updated version that is in alignment with labeling requirements specified in the IEC-60101-1 standard.

    FDA · DeviceDeVilbiss Healthcare LLCWorldwide - US Nationwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls VITROS 5600 Integrated System- For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a…

    MicroTip Pack Opener Assembly Potentially Not Removing or Replacing MicroTip Pack Caps on VITROS¿ 5600 and XT 7600 Integrated Systems, results may be delayed results if a pack becomes unusable

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Nationwide including Puerto Rico and Bermuda Foreign: Australia Belgium Brazil…Official notice →
  5. US
    Drugs & medicalClass II

    DermLite LLC recalls DERMLITE DL4W, REF: DL4W, Rx Only MD, CE

    Incorrect labeling; Package labeling contains a different serial number then the serial number on the product labeling.

    FDA · DeviceDermLite LLCWorldwide - US Nationwide distribution in the states of CA, AZ, TX, FL, NE,…Official notice →
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Cortisol 50T (Material Number 10995538), 250T (Material Number 10995537), and REF 250T (Material Number…

    There is a negative bias with urine patient samples and urine cortisol Quality Control (QC) results intermittently out of range low.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide.Official notice →
  7. US
    Drugs & medicalClass II

    Spacelabs Healthcare, Inc. recalls Ultraview SL (UVSL) Command Module, Model 91496

    Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent…

    FDA · DeviceSpacelabs Healthcare, Inc.US Nationwide distribution in the states of SD, KY, MO, WA, OH, NC, CA, OR,…Official notice →
See all drugs & medical recalls →