Drugs & medicalClass II

Ortho-Clinical Diagnostics recalls VITROS Performance Verifier I

The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing…

Read the official notice →
Date
2024-04-03
Country
United States
Brand
Ortho-Clinical Diagnostics
Product
VITROS Performance Verifier I
Classification
Class II
Units
Worldwide distribution - US Nationwide and the countries of Australia,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Ortho-Clinical Diagnostics recalls

  1. 2024-07-10
  2. US
    Drugs & medicalClass III

    Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS…

    Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration date for Lot 2953 is 24-Jul-2024. If used past its true expiration and the customer does not have their Assay Data Disk…

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide including Puerto Rico and the countries…Official notice →
  3. 2024-06-19
  4. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Anti HBs Calibrators. The VITROS Anti-HBs Calibrator Pack contains 3 X labelled Calibrator…

    Potential for positively biased results due to signal reduction over shelf life.

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →
  5. 2024-06-12
  6. US
    Drugs & medicalClass III

    Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Vitamin B12 Range Verifier - In vitro diagnostic use ly. For use in verifying the calibration…

    The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 1200 RV Level 2 may cause range verification failure. The issue is detectable by the user when the product fails to meet the…

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide and the country of Columbia.Official notice →
  7. 2024-03-13
  8. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls VITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (refurbished)

    A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA,…Official notice →
  9. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls VITROS 5600 Integrated System (New and Refurbished)

    Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VITROS¿ 3600 Immunodiagnostic Systems and VITROS¿ 5600/XT 7600 Integrated Systems which may lead to erroneous test results.

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  10. 2024-01-31
  11. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Anti-HBs Quantitative Reagent Pack, Catalog No. 6801925 VITROS Immunodiagnostic Products…

    Affected lots may experience increased calibration failures or an increase in falsely elevated results.

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Domestic distribution nationwide. OUS distribution to Bermuda, Australia,…Official notice →
All 44 Ortho-Clinical Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2024-04-03
  2. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Substrate, REF 81906, For use with the Access Immunoassay Systems

    On 07 February 2024 Beckman Coulter made the decision to conduct a field action on Access Substrate (81906) lots: 234600, 337984, 338040, 338214, 338229, 338289, 338322, 338333, 338451, 338587, and 338538. Beckman…

    FDA · DeviceBeckman Coulter, Inc.Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of:…Official notice →
  3. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Insufflation unit, UHI-4.

    Olympus has become aware of an increased trend of both repairs and customer complaints reporting "UHI-4 stopping air supply due to the front panel LED turning off".

    FDA · DeviceOlympus Corporation of the AmericasDomestic: Nationwide Distribution.Official notice →
  4. US
    Drugs & medicalClass I

    Boston Scientific Corporation recalls Obsidio Conformable Embolic, Material Numbers (UPN): a) M0013972001010, b) M0013972101010

    An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common…

    FDA · DeviceBoston Scientific CorporationUSOfficial notice →
  5. US
    Drugs & medicalClass II

    Hologic, Inc. recalls Aptima CMV Quant Assay, Reference numbers: AW-22600-001, AW-25509-001, and AW-26381-001 Amplification Reagent, Enzyme Reagent, Promoter…

    Use of assay along with plasma specimen may result in error code which would invalidate sample.

    FDA · DeviceHologic, Inc.Worldwide - US Nationwide distribution in the states of CA, CT, FL, ID, IL,…Official notice →
  6. US
    Drugs & medicalClass II

    Alcon Research, LLC recalls SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs, Products: 18018-06, 3865-44,…

    Medline Bulb irrigation syringes recalled due to potential for packaging sterile barrier to be breached, and these syringes were packed inside surgical procedure packs that were subsequently terminally sterilized, so…

    FDA · DeviceAlcon Research, LLCUS Nationwide distribution in the states of OH, WV.Official notice →
  7. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Cystatin C Calibrator - In vitro diagnostic (IVD) use in the calibration of the Randox Cystatin C assay on automated analysers Catalog…

    Cystatin C Calibrator Series CVS 2699, is running with a negative bias compared to other methods

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of CA, IN.Official notice →
See all drugs & medical recalls →