Abbott Laboratories recalls Alinity hq Analyzer, REF: 09P68-01
1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating…
Read the official notice →- Date
- 2024-08-28
- Country
- United States
- Brand
- Abbott Laboratories
- Product
- Alinity hq Analyzer, REF: 09P68-01
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of NY, OK, NC and the…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Abbott Laboratories recalls
- 2025-12-24
US
Abbott Laboratories recalls Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20
Abbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction. The reason for the recall is potential performance issues found in the Alinity ci-series…
FDA · DeviceAbbott LaboratoriesWorldwide distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO,…Official notice →- 2025-11-19
US
Abbott Laboratories recalls Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version: 5.0 to 5.10.1 for OUS,…
Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.
FDA · DeviceAbbott LaboratoriesWorldwide distribution: US (nationwide) to states of: AL, CA, CO, FL, IN, MN,…Official notice →- 2025-04-30
US
Abbott Laboratories recalls Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01 Software Version: All software versions Product…
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is…
FDA · DeviceAbbott LaboratoriesWorldwide distribution: US (nationwide) to the following states: AR, AZ, CA,…Official notice →- 2024-04-10
US
Abbott Laboratories recalls ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.
The reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may lead to incorrect myoglobin results and delayed diagnosis…
FDA · DeviceAbbott LaboratoriesWorldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL,…Official notice →- 2024-01-17
US
Abbott Laboratories recalls Alinity i Processing Module REF 03R65-01 The Alinity i Processing Module is a fully automated immunoassay analyzer allowing random and…
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
FDA · DeviceAbbott LaboratoriesU.S.: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA,…Official notice →- 2024-01-03
US
Abbott Laboratories recalls Alinity ci-series System Control Module, REF: 03R70-01
There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower, including: 1) SCC Reagent load error during daily maintenance; 2) Sample Laboratory Report error; 3)…
FDA · DeviceAbbott LaboratoriesUS: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA,…Official notice →
Other drugs & medical recalls in the United States
- 2024-08-28
US
ROi CPS LLC recalls Regard Kit containing the Microtek decanter
Decanters may have pin holes and tears in sterile packaging that cannot be identified visually.
US
Ortho-Clinical Diagnostics, Inc. recalls The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and a total protein test…
The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Nationwide. Foreign: Australia Belgium Bermuda Brazil Canada L3R 0B8 Chile…Official notice →US
BioFire Diagnostics, LLC recalls BIOFIRE FILMARRAY Pneumonia Panel plus, REF: RFIT-ASY-0143
Pneumonia panels have an increased risk of control failures due to an issue in the manufacturing process, which could lead to delayed results for the customer.
FDA · DeviceBioFire Diagnostics, LLCWorldwide - US Nationwide distribution in the states of WV, IN, AL and the…Official notice →US
Alcon Research LLC recalls Product Desc: Product Name: CLEARCUT SB INTREPID 2.2 ANG Model/Catalog Number: 8065992561 Software Version: N/A Product Description:…
Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives.
FDA · DeviceAlcon Research LLCDomestic: Nationwide Distribution Countries Affected: Argentina, Australia,…Official notice →US
Ventana Medical Systems, Inc. recalls Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody, REF: 790-4506, IVD
Their is a potential for high background and off-target staining when using the monoclonal primary antibody which can result in false positive results.
FDA · DeviceVentana Medical Systems, Inc.Worldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →US
Microtek Medical Inc. recalls Bag Decanter II, REF 2002S. Packaged on a form, fill and seal machine with a poly/Tyvek material then placed 50 products to a case (cs).
During sterile barrier testing performed on the decanter product line, the samples were identified to have pin holes and tears in the packaging film.
FDA · DeviceMicrotek Medical Inc.Distributed nationwide and internationally to Canada, Mexico, New Zealand,…Official notice →