Drugs & medicalClass II

Abbott Laboratories recalls Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version: 5.0 to 5.10.1 for OUS,…

Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.

Read the official notice →
Date
2025-11-19
Country
United States
Brand
Abbott Laboratories
Product
Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version: 5.0 to 5.10.1 for OUS,…
Classification
Class II
Units
Worldwide distribution: US (nationwide) to states of: AL, CA, CO, FL, IN, MN,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abbott Laboratories recalls

  1. 2025-12-24
  2. US
    Drugs & medicalClass II

    Abbott Laboratories recalls Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20

    Abbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction. The reason for the recall is potential performance issues found in the Alinity ci-series…

    FDA · DeviceAbbott LaboratoriesWorldwide distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO,…Official notice →
  3. 2025-04-30
  4. US
    Drugs & medicalClass II

    Abbott Laboratories recalls Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01 Software Version: All software versions Product…

    When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is…

    FDA · DeviceAbbott LaboratoriesWorldwide distribution: US (nationwide) to the following states: AR, AZ, CA,…Official notice →
  5. 2024-08-28
  6. US
    Drugs & medicalClass II

    Abbott Laboratories recalls Alinity hq Analyzer, REF: 09P68-01

    1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating…

    FDA · DeviceAbbott LaboratoriesWorldwide - US Nationwide distribution in the states of NY, OK, NC and the…Official notice →
  7. 2024-04-10
  8. US
    Drugs & medicalClass II

    Abbott Laboratories recalls ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.

    The reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may lead to incorrect myoglobin results and delayed diagnosis…

    FDA · DeviceAbbott LaboratoriesWorldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL,…Official notice →
  9. 2024-01-17
  10. US
    Drugs & medicalClass II

    Abbott Laboratories recalls Alinity i Processing Module REF 03R65-01 The Alinity i Processing Module is a fully automated immunoassay analyzer allowing random and…

    The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).

    FDA · DeviceAbbott LaboratoriesU.S.: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA,…Official notice →
  11. 2024-01-03
  12. US
    Drugs & medicalClass II

    Abbott Laboratories recalls Alinity ci-series System Control Module, REF: 03R70-01

    There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower, including: 1) SCC Reagent load error during daily maintenance; 2) Sample Laboratory Report error; 3)…

    FDA · DeviceAbbott LaboratoriesUS: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA,…Official notice →
All 21 Abbott Laboratories recalls →

Other drugs & medical recalls in the United States

  1. 2025-11-19
  2. US
    Drugs & medicalClass II

    Applied Medical Technology Inc recalls Applied Medical Technology, Inc., NutraGlide, 6 F- 8 F x 90 Nasal Feeding Tube with Stylet and ENfit, REF NFT-06090-I

    The nasal feeding tube may be occluded

    FDA · DeviceApplied Medical Technology IncUS Nationwide distribution in the states of MA, RI, NC, OH.Official notice →
  3. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;

    The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution in the state of CA and the countries of…Official notice →
  4. US
    Drugs & medicalClass II

    Stryker Sustainability Solutions recalls HARMONIC FOCUS Shears + Adaptive Tissue Technology, 9 cm Length REF: HAR9F, Reprocessed Device for Single Use, STERILEEO, RX ONLY

    Due to manufacturing issue, reprocessed shears, may become stuck in the "on position" resulting in continuous activation.

    FDA · DeviceStryker Sustainability SolutionsUS Nationwide distribution and OUS (foreign) to country of: Canada.Official notice →
  5. US
    Drugs & medicalClass II

    Medtronic Perfusion Systems recalls Affinity NT Oxygenator, Oxygenator with Balance Biosurface and Uncoated Cardiotomy/Venous Reservoir, Product Number 541B

    Medtronic identified an Affinity NT Oxygenator unit with decreased CO2 gas transfer rates that did not meet performance requirements.

    FDA · DeviceMedtronic Perfusion SystemsUS States: AZ, CA, TNOfficial notice →
  6. US
    Drugs & medicalClass II

    Covidien, LP recalls Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;

    A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.

    FDA · DeviceCovidien, LPWorldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →
  7. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Kiestra" Isolate Suspension Cuvette Array; Catalog No.: 246100;

    The 2D barcodes on these cuvettes contain formatting errors that prevent the analyzing system from correctly reading the expiration date. As a result, the instrument enters an error state and the cuvettes cannot be used.

    FDA · DeviceBecton Dickinson & Co.Worldwide - US Nationwide distribution in the states of CO, CT, KY, NJ, TN,…Official notice →
See all drugs & medical recalls →