Siemens Healthcare Diagnostics, Inc. recalls Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode…

Read the official notice →
Date
2025-11-19
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Classification
Class II
Units
Worldwide - US Nationwide distribution in the state of CA and the countries of…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2026-05-06
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.

    Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  3. 2026-04-08
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;

    A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Domestic: Nationwide Distribution International: Argentina, Australia,…Official notice →
  5. 2026-03-11
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urine Albumin (UAlb). Material Number: 11537225

    Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  7. 2025-11-05
  8. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number:…

    The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of AL, CA, DC, FL, ID,…Official notice →
  9. 2025-09-17
  10. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP). Material Number: 11097543.

    Potential for falsely depressed patient, quality control (QC), and/or calibration results or delayed results.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  11. 2025-08-13
  12. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.

    Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L.…

    FDA · DeviceSiemens Healthcare Diagnostics IncWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2025-11-19
  2. US
    Drugs & medicalClass III

    Accriva Diagnostics, Inc. recalls VerifyNow PRUTest Platelet Reactivity Test REF 85225 UDI-DI code: 10711234150078 The VerifyNow PRUTest is a whole blood test used in…

    Due to an device without a premarket clearance being incorrectly package and distributed.

    FDA · DeviceAccriva Diagnostics, Inc.Worldwide - U.S. Nationwide distribution in the states of AK, CA, FL, IA, IL,…Official notice →
  3. US
    Drugs & medicalClass II

    SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION recalls Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems…

    Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the rotating shaft.

    FDA · DeviceSHIMADZU CORPORATION MEDICAL SYSTEMS DIVISIONWorldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →
  4. US
    Drugs & medicalClass II

    Applied Medical Technology Inc recalls Applied Medical Technology, Inc., NutraGlide, 6 F- 8 F x 90 Nasal Feeding Tube with Stylet and ENfit, REF NFT-06090-I

    The nasal feeding tube may be occluded

    FDA · DeviceApplied Medical Technology IncUS Nationwide distribution in the states of MA, RI, NC, OH.Official notice →
  5. US
    Drugs & medicalClass II

    Stryker Sustainability Solutions recalls HARMONIC FOCUS Shears + Adaptive Tissue Technology, 9 cm Length REF: HAR9F, Reprocessed Device for Single Use, STERILEEO, RX ONLY

    Due to manufacturing issue, reprocessed shears, may become stuck in the "on position" resulting in continuous activation.

    FDA · DeviceStryker Sustainability SolutionsUS Nationwide distribution and OUS (foreign) to country of: Canada.Official notice →
  6. US
    Drugs & medicalClass II

    Abbott Laboratories recalls Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version: 5.0 to 5.10.1 for OUS,…

    Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.

    FDA · DeviceAbbott LaboratoriesWorldwide distribution: US (nationwide) to states of: AL, CA, CO, FL, IN, MN,…Official notice →
  7. US
    Drugs & medicalClass II

    Medtronic Perfusion Systems recalls Affinity NT Oxygenator, Oxygenator with Balance Biosurface and Uncoated Cardiotomy/Venous Reservoir, Product Number 541B

    Medtronic identified an Affinity NT Oxygenator unit with decreased CO2 gas transfer rates that did not meet performance requirements.

    FDA · DeviceMedtronic Perfusion SystemsUS States: AZ, CA, TNOfficial notice →