Drugs & medicalClass II

Siemens Healthcare Diagnostics, Inc. recalls Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number:…

The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on…

Read the official notice →
Date
2025-11-05
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2,…
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of AL, CA, DC, FL, ID,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2026-04-08
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;

    A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Domestic: Nationwide Distribution International: Argentina, Australia,…Official notice →
  3. 2026-03-11
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urine Albumin (UAlb). Material Number: 11537225

    Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  5. 2025-11-19
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;

    The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution in the state of CA and the countries of…Official notice →
  7. 2025-09-17
  8. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP). Material Number: 11097543.

    Potential for falsely depressed patient, quality control (QC), and/or calibration results or delayed results.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  9. 2025-08-13
  10. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.

    Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L.…

    FDA · DeviceSiemens Healthcare Diagnostics IncWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  11. 2025-08-06
  12. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

    Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride,…

    FDA · DeviceSiemens Healthcare Diagnostics Incworldwide distribution - US Nationwide and the countries of Canada, Chile,…Official notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2025-11-05
  2. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS recalls Model: Incisive CT for Brazil SKD; Product Code (REF): 728146; Software Versions: 5.1.0.X & 5.1.1.X;

    Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/…

    FDA · DevicePHILIPS MEDICAL SYSTEMSDomestic: AL, AR, CA, CT, FL, GA, IL, IN, KY, MD, MI, OH, PA, PR, TN, TX, UT,…Official notice →
  3. US
    Drugs & medicalClass II

    ASP Global, LLC. dba Anatomy Supply Partners, LLC. recalls SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011;…

    All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False…

    FDA · DeviceASP Global, LLC. dba Anatomy Supply Partners, LLC.Worldwide - US Nationwide distribution including in the states of AK, AL, AZ,…Official notice →
  4. US
    Drugs & medicalClass II

    CareFusion 303, Inc. recalls BD Pyxis Pro MedStation Main, REF: 1155-00

    Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.

    FDA · DeviceCareFusion 303, Inc.Worldwide distribution - US Nationwide and the countries of Canada, United…Official notice →
  5. US
    Drugs & medicalClass II

    ICU Medical, Inc. recalls ICU Medical ChemoLock Closed Vial Spike w/Skirt, 5 Units REF: CL-80S-5

    Port weld of drug transfer device may separate or break during use and potentially result in a leak.

    FDA · DeviceICU Medical, Inc.Worldwide distribution: US (Nationwide) including states of: AK AL AR AZ CA CO…Official notice →
  6. US
    Drugs & medicalClass II

    Canon Medical System, USA, INC. recalls VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.

    Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.

    FDA · DeviceCanon Medical System, USA, INC.US distribution to Ohio and NevadaOfficial notice →
  7. US
    Drugs & medicalClass II

    Alcon Research LLC recalls Alcon PRECISION1 One-Day Contact Lenses, -2.25D.

    Potential for a weak seal in some units resulting in compromise in sterility.

    FDA · DeviceAlcon Research LLCWorldwide - US distribution in the states of Alabama, Arizona, Arkansas,…Official notice →
See all drugs & medical recalls →