Beckman Coulter Mishima K.K. recalls Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965…
Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC…
Read the official notice →- Date
- 2026-10-07
- Country
- United States
- Brand
- Beckman Coulter Mishima K.K
- Product
- Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693…
- Classification
- Class II
- Units
- Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Beckman Coulter Mishima K.K recalls
- 2026-03-18
US
Beckman Coulter Mishima K.K. recalls CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases…
FDA · DeviceBeckman Coulter Mishima K.K.US Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC,…Official notice →- 2025-06-11
US
Beckman Coulter Mishima K.K. recalls Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 -…
Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of…
FDA · DeviceBeckman Coulter Mishima K.K.Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX,…Official notice →US
Beckman Coulter Mishima K.K. recalls Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 -…
A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing.…
FDA · DeviceBeckman Coulter Mishima K.K.Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX,…Official notice →- 2024-08-28
US
Beckman Coulter Mishima K.K. recalls DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in…
Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will…
FDA · DeviceBeckman Coulter Mishima K.K.Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL,…Official notice →- 2024-08-07
US
Beckman Coulter Mishima K.K. recalls DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in…
There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient…
FDA · DeviceBeckman Coulter Mishima K.K.Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL,…Official notice →
Other drugs & medical recalls in the United States
- 2026-10-07
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;
The potential for falsely elevated results.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →US
Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System
Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the…
FDA · DeviceWright Medical Technology, Inc.US Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN,…Official notice →US
Karl Storz Endoscopy recalls HOPKINS Telescope 30¿, 4mm REF 10005BA UDI-DI code: 04048551045121 KARL STORZ HOPKINS¿ and HOPKINS¿ II telescopes are to be used for…
Due to incorrect inside packaging label for eyepiece.
FDA · DeviceKarl Storz EndoscopyU.S. Nationwide distribution in the states of AZ, CA and MO.Official notice →US
Penumbra Inc. recalls Brand Name: Penumbra System (RED 62 Reperfusion Catheter) Product Name: Penumbra System (RED 62 Reperfusion Catheter) Model/Catalog…
Due to manufacturing issues reperfusion catheters may fracture.
FDA · DevicePenumbra Inc.US Nationwide distribution in the states of AL, CA, CO, FL, LA, NC, NY, TX,…Official notice →US
Tandem Diabetes Care, Inc. recalls Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version…
A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous…
US
Olympus Corporation of the Americas recalls Olympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunction wtih OLYSENSE…
While in the high-power operating mode, a video dropout/blackout may occur when using a particular accessory cable.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the country of Poland.Official notice →