Drugs & medicalClass II

Tandem Diabetes Care, Inc. recalls Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version…

A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous…

Read the official notice →
Date
2026-10-07
Country
United States
Brand
Tandem Diabetes Care
Product
Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13…
Classification
Class II
Units
US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Tandem Diabetes Care recalls

  1. 2026-04-01
  2. US
    Drugs & medicalClass II

    Tandem Diabetes Care, Inc. recalls Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven…

    The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead…

    FDA · DeviceTandem Diabetes Care, Inc.International distribution in the countries of Czech Republic, Slovakia.Official notice →
  3. 2025-11-12
  4. US
    Drugs & medicalClass I

    Tandem Diabetes Care, Inc. recalls Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and…

    Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot…

    FDA · DeviceTandem Diabetes Care, Inc.US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY,…Official notice →
  5. 2025-10-29
  6. US
    Drugs & medicalClass II

    Tandem Diabetes Care, Inc. recalls t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide…

    A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an…

    FDA · DeviceTandem Diabetes Care, Inc.US nationwide distribution including Puerto Rico.Official notice →
  7. 2025-10-01
  8. US
    Drugs & medicalClass II

    Tandem Diabetes Care, Inc. recalls t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)

    an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent…

    FDA · DeviceTandem Diabetes Care, Inc.International distribution to the country of Israel.Official notice →
  9. 2025-09-03
  10. US
    Drugs & medicalClass I

    Tandem Diabetes Care, Inc. recalls Battery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished,…

    Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in…

    FDA · DeviceTandem Diabetes Care, Inc.Worldwide distribution. US nationwide including Puerto Rico, Canada, Czech…Official notice →
  11. 2025-04-23
  12. US
    Drugs & medicalClass I

    Tandem Diabetes Care, Inc. recalls t:slim X2 Insulin Pump with Interoperable Technology

    A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated…

    FDA · DeviceTandem Diabetes Care, Inc.US: AZ, CA, CO, CT, FL, HI, IA, ID, IL, IN, MD, ME, MI, MO, MS, NC, ND, NE,…Official notice →
All 10 Tandem Diabetes Care recalls →

Other drugs & medical recalls in the United States

  1. 2026-10-07
  2. US
    Drugs & medicalClass II

    Karl Storz Endoscopy recalls HOPKINS Telescope 30¿, 4mm REF 10005BA UDI-DI code: 04048551045121 KARL STORZ HOPKINS¿ and HOPKINS¿ II telescopes are to be used for…

    Due to incorrect inside packaging label for eyepiece.

    FDA · DeviceKarl Storz EndoscopyU.S. Nationwide distribution in the states of AZ, CA and MO.Official notice →
  3. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965…

    Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA,…Official notice →
  4. US
    Drugs & medicalClass II

    Penumbra Inc. recalls Brand Name: Penumbra System (RED 62 Reperfusion Catheter) Product Name: Penumbra System (RED 62 Reperfusion Catheter) Model/Catalog…

    Due to manufacturing issues reperfusion catheters may fracture.

    FDA · DevicePenumbra Inc.US Nationwide distribution in the states of AL, CA, CO, FL, LA, NC, NY, TX,…Official notice →
  5. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Olympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunction wtih OLYSENSE…

    While in the high-power operating mode, a video dropout/blackout may occur when using a particular accessory cable.

    FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the country of Poland.Official notice →
  6. US
    Drugs & medicalClass II

    SPECTRUM MEDICAL SRL recalls Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 W Model/Catalog Number: HX55V-S0W…

    Due to blood-to-glycol leak in the heat exchanger.

    FDA · DeviceSPECTRUM MEDICAL SRLUS Nationwide distribution in the states of AL, CA, FL, KS MA, MD, MI, MO, NC,…Official notice →
  7. US
    Drugs & medicalClass II

    Micro Therapeutics, Inc. recalls Pipeline Flex Embolization Device with Shield Technology REF PED2-300-12 UDI-DI code: 00763000448806 he Pipeline Flex Embolization…

    Due to incorrect label/product dimensions. Product is supposed to be 3.00 mm in diameter, but unit is packaging was a 2.50 mm in diameter.

    FDA · DeviceMicro Therapeutics, Inc.U.S. Nationwide distribution in the states of GA and TX.Official notice →
See all drugs & medical recalls →