Tandem Diabetes Care, Inc. recalls t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide…
A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an…
Read the official notice →- Date
- 2025-10-29
- Country
- United States
- Brand
- Tandem Diabetes Care
- Product
- t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when…
- Classification
- Class II
- Units
- US nationwide distribution including Puerto Rico.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Tandem Diabetes Care recalls
- 2026-10-07
US
Tandem Diabetes Care, Inc. recalls Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version…
A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous…
- 2026-04-01
US
Tandem Diabetes Care, Inc. recalls Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven…
The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead…
FDA · DeviceTandem Diabetes Care, Inc.International distribution in the countries of Czech Republic, Slovakia.Official notice →- 2025-11-12
US
Tandem Diabetes Care, Inc. recalls Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and…
Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot…
FDA · DeviceTandem Diabetes Care, Inc.US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY,…Official notice →- 2025-10-01
US
Tandem Diabetes Care, Inc. recalls t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)
an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent…
FDA · DeviceTandem Diabetes Care, Inc.International distribution to the country of Israel.Official notice →- 2025-09-03
US
Tandem Diabetes Care, Inc. recalls Battery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished,…
Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in…
FDA · DeviceTandem Diabetes Care, Inc.Worldwide distribution. US nationwide including Puerto Rico, Canada, Czech…Official notice →- 2025-04-23
US
Tandem Diabetes Care, Inc. recalls t:slim X2 Insulin Pump with Interoperable Technology
A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated…
FDA · DeviceTandem Diabetes Care, Inc.US: AZ, CA, CO, CT, FL, HI, IA, ID, IL, IN, MD, ME, MI, MO, MS, NC, ND, NE,…Official notice →
Other drugs & medical recalls in the United States
- 2025-10-29
US
Becton Dickinson & Co. recalls BD EpiCenter Microbiology Data Management System, labeled as the following: 1. Computer System Epicenter F/G, Catalog Number: 441002. 2.…
Product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of…
FDA · DeviceBecton Dickinson & Co.Worldwide distribution. US Nationwide including Guam and Puerto Rico, Algeria,…Official notice →US
Medline Industries, LP recalls Excelsior Medical, 0.9% Sodium Chloride Injection, USP - 10 mL Fill, Item Number EMZ111240. Prefilled saline flush syringes
Medline has identified a limited quantity of saline IV flush syringes that may not meet the required concentration of sodium chloride.
US
Baxter Healthcare Corporation recalls Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 01690-400)…
Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing…
FDA · DeviceBaxter Healthcare CorporationDistribution US nationwide, Canada and Colombia.Official notice →US
GE Medical Systems, LLC recalls MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001,…
If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires.…
FDA · DeviceGE Medical Systems, LLCWorldwide Distribution: US (nationwide) and OUS (foreign/Global) to countries…Official notice →US
CPM Medical Consultants, LLC. recalls Fuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L, OS200025L, OS200120L,…
Instructions for use for a foot plating system is being updated to remove the second option of requirements for sterilization which was 3 minutes exposure time at 134-137-degrees C (273-279-degrees F) and 40mBar for…
FDA · DeviceCPM Medical Consultants, LLC.US Nationwide distribution in the states of NV, TX, LA, WA, WI, FL.Official notice →US
Synthes (USA) Products LLC recalls LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral…
Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as…
FDA · DeviceSynthes (USA) Products LLCWorldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL,…Official notice →