VIRTUAL INCISION CORPORATION recalls MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version…

The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the…

Read the official notice →
Date
2025-06-11
Country
United States
Brand
VIRTUAL INCISION CORPORATION
Product
MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3
Classification
Class II
Units
US Distribution to States: CA, FL, NE
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

Other drugs & medical recalls in the United States

  1. 2025-06-11
  2. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 -…

    Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX,…Official notice →
  3. US
    Drugs & medicalClass II

    Beckman Coulter Mishima K.K. recalls Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 -…

    A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing.…

    FDA · DeviceBeckman Coulter Mishima K.K.Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX,…Official notice →
  4. US
    Drugs & medicalClass II

    Dexcom, Inc. recalls Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system

    A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are…

    FDA · DeviceDexcom, Inc.Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of:…Official notice →
  5. US
    Drugs & medicalClass II

    Luminex Corporation recalls VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System;…

    It was determined that the consumable issue may result in No Call - INT CTL 1 and/or False Negative results on a small portion of the lot of test cartridges.

    FDA · DeviceLuminex CorporationUS Nationwide distribution in the states of AL, CA, MN, TX, AZ, NJ, WA, SC,…Official notice →
  6. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Testosterone assay, Catalog Number 33560

    Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result…

    FDA · DeviceBeckman Coulter, Inc.US State: FL. UAE, VietnamOfficial notice →
  7. US
    Drugs & medicalClass II

    Silex Medical, LLC recalls Signature Laparoscopic Instruments, Bullet Tip Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF…

    Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.

    FDA · DeviceSilex Medical, LLCUS Nationwide distribution in the states of Florida, North Dakota, Illinois,…Official notice →