Luminex Corporation recalls VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System;…
It was determined that the consumable issue may result in No Call - INT CTL 1 and/or False Negative results on a small portion of the lot of test cartridges.
Read the official notice →- Date
- 2025-06-11
- Country
- United States
- Brand
- Luminex
- Product
- VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System;…
- Classification
- Class II
- Units
- US Nationwide distribution in the states of AL, CA, MN, TX, AZ, NJ, WA, SC,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Luminex recalls
- 2025-04-25
PL
EU
Fixed lamp (LUMINEX) · RENO
A live wire is in contact with the metal casing of the lamp. The user could touch accessible live parts and receive an electric shock. The product does not comply with the requirements of the Low Voltage Directive, nor…
- 2024-08-14
US
Luminex Corporation recalls Verigene Gram Positive Blood Culture (BC-GP) Nucleic Acid Test
Potential for defective test cartridges which may result in false positive results for the Staphylococcus lugdunensis and Enterococcus faecium targets.
FDA · DeviceLuminex CorporationUS Nationwide distribution including in the states of AZ, CA, CO, DE, FL, GA,…Official notice →- 2023-12-13
US
Luminex Corporation recalls Luminex VERIGENE BC-GP Utility Tray (5-Pack Carrier), Gram-Positive Blood Culture Nucleic Acid Test, REF 20-011-018, a component of the…
Potentially defective utility trays in the reagent kits.
FDA · DeviceLuminex CorporationWorldwide - US Nationwide distribution in the states of AZ, CA, CO, DE, FL,…Official notice →- 2023-07-05
US
Luminex Corporation recalls VERIGENE Clostridium difficile Nucleic Acid Test, REF 30-002-022
It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a false-negative result…
- 2022-11-02
US
Luminex Corporation recalls ARIES SARS-CoV-2 Assay, REF: 50-10047
A part defect with the wash 1 syringe of an SARS-CoV-2 assay could lead to incomplete extraction thereby producing an invalid or false negative assay result.
FDA · DeviceLuminex CorporationDistribution US nationwide, Indonesia, and Hong Kong.Official notice →- 2022-08-24
US
Luminex Corporation recalls ARIES SARS-CoV-2 Assay
SARS-CoV-2 assay that detects SARS-CoV-2 ORF1ab and N genes may not detect the ORF1ab gene. A positive results if either SARS-CoV-2 genes is detected, but If both genes are mistakenly not detected, a false negative…
FDA · DeviceLuminex CorporationU.S. Nationwide distribution in the states of NE, TX, MS, TN, KY, ME, CA.Official notice →
Other drugs & medical recalls in the United States
- 2025-06-11
US
Beckman Coulter Mishima K.K. recalls Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 -…
A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing.…
FDA · DeviceBeckman Coulter Mishima K.K.Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX,…Official notice →US
Dexcom, Inc. recalls Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system
A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are…
FDA · DeviceDexcom, Inc.Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of:…Official notice →US
VIRTUAL INCISION CORPORATION recalls MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version…
The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the…
US
Beckman Coulter, Inc. recalls Access Testosterone assay, Catalog Number 33560
Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result…
US
Silex Medical, LLC recalls Signature Laparoscopic Instruments, Bullet Tip Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF…
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
FDA · DeviceSilex Medical, LLCUS Nationwide distribution in the states of Florida, North Dakota, Illinois,…Official notice →US
Civco Medical Instruments Co. Inc. recalls BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic…
During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.
FDA · DeviceCivco Medical Instruments Co. Inc.OUS Distribution to countries of: Japan, Taiwan, France, Georgia, Germany and…Official notice →