Siemens Healthcare Diagnostics, Inc. recalls Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and…
A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or…
Read the official notice →- Date
- 2024-11-27
- Country
- United States
- Brand
- Siemens Healthcare Diagnostics
- Product
- Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and present them to…
- Classification
- Class II
- Units
- US Nationwide and Worldwide distribution: Foreign: Algeria Algeria Argentina…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Healthcare Diagnostics recalls
- 2025-01-01
US
Siemens Healthcare Diagnostics, Inc. recalls epoc NXS Host; Siemens Material Number (SMN): 11413498(JP) ; Software Version Number: 4.14.9. Intended for use by trained…
Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain…Official notice →- 2024-12-11
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in…
Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH ¿2-Microglobulin (B2M) reagent.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →US
Siemens Healthcare Diagnostics, Inc. recalls Atellica CI Analyzer. Catalog Numbers: 10947347.
Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Austria,…Official notice →- 2024-11-20
US
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)
The IMMULITE 2000 PSA assay for use on the IMMULITE 2000 and IMMULITE 2000 XPi analyzers did not meet the High-dose Hook Effect claim as stated in the Instructions for Use (IFU).
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution.Official notice →- 2024-11-13
US
Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from…
- 2024-10-09
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM aTPO Assay, Material Number 10995466 (100 Test) and 10995467 (500 Test)
Positive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide.Official notice →
Other drugs & medical recalls in the United States
- 2024-11-27
US
Cook Incorporated recalls Approach CTO Micro Wire Guide, RPN CMW-14-190-25G, GPN G50789
Devices from the affected device lots have labels that state the incorrect expiration dates.
US
Olympus Corporation of the Americas recalls Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number:…
To better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors…
FDA · DeviceOlympus Corporation of the AmericasDomestic: US Nationwide Distribution.Official notice →US
Philips Medical Systems Technologies Ltd. recalls Vue PACS Diagnostic Viewer with Vue PACS client 12.1.5 to 12.2.8.454 software Model Number: 1149418 Intended use is to…
A software issue affecting Philips Vue PACS may cause incorrect ischemic map and table value calculations when using the export function in the Perfusion Application for a Siemens CT perfusion study, resulting in…
FDA · DevicePhilips Medical Systems Technologies Ltd.US Nationwide and Worldwide Distribution Foreign: Country Argentina Australia…Official notice →US
Abbott Medical recalls Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile
A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or…
FDA · DeviceAbbott MedicalNew York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman,…Official notice →US
Alcon Research LLC recalls a) Alcon Custom-Pak CONSTELLATION 25G PLUS CHRISTIANA CARE HLTH SERV, # 15340-11, containing TUBING, SALEM SUMP, 14 FR; b) Alcon…
Reported breakage in the tubing Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.
FDA · DeviceAlcon Research LLCUS distribution to states of: Delaware, Illinois, and OklahomaOfficial notice →US
Roche Diagnostics Operations, Inc. recalls cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total…
Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)
FDA · DeviceRoche Diagnostics Operations, Inc.US: NJ, TX, NY, FL, IN, TX, OK, SC, IA,Official notice →