Drugs & medicalClass II

Abbott Medical recalls Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile

A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or…

Read the official notice →
Date
2024-11-27
Country
United States
Brand
Abbott Medical
Product
Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile
Classification
Class II
Units
New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abbott Medical recalls

  1. 2025-04-23
  2. US
    Drugs & medicalClass II

    Abbott Medical recalls Aveir DR Leadless Pacemaker (LP) Ventricular Device REF LSP202V UDI-DI code: 05415067040725 REF LSP112V UDI-DI code: 05415067034472 The…

    Due to a manufacturing issue, leadless pacemakers may have been exposed to higher than normal temperatures which may result in premature battery failure.

    FDA · DeviceAbbott MedicalWorldwide - U.S. Nationwide distribution in the states of FL, SC, and WV and…Official notice →
  3. 2025-01-01
  4. US
    Drugs & medicalClass II

    Abbott Medical recalls The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates…

    Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the…

    FDA · DeviceAbbott MedicalWorldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →
  5. 2024-12-04
  6. US
    Drugs & medicalClass III

    Abbott Medical recalls Merlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data from the patient s implanted…

    Due to heart transmitter not being able to upgrade to the current software version in the field after experiencing a power interruption during a prior over-the-wire software download.

    FDA · DeviceAbbott MedicalWorldwide - U.S. Nationwide distribution in the states of ALABAMA, ALASKA,…Official notice →
  7. 2024-07-31
  8. US
    Drugs & medicalClass II

    Abbott Medical recalls Abbott Infinity 7, Implantable Pulse Generator, REF: 6663, SterileEO, Rx Only

    Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.

    FDA · DeviceAbbott MedicalWorldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →
  9. 2024-07-24
  10. US
    Drugs & medicalClass II

    Abbott Medical recalls Abbott Proclaim XR 5 Implantable Pulse Generator, REF 3660, Sterile EO, Rx Only

    The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.

    FDA · DeviceAbbott MedicalWorldwide - US Nationwide distribution including in the states of ALABAMA,…Official notice →
  11. 2024-07-10
  12. US
    Drugs & medicalClass II

    Abbott Medical recalls Abbott Liberta RC, Implantable Pulse Generator, REF: 62400, STERILE, RxONLY

    Deep brain stimulation system will first turn off after approximately 50 days after the system is activated and subsequently, this same action will recur approximately every 50 days thereafter. This unexpected lost of…

    FDA · DeviceAbbott MedicalUS: Nationwide distribution including in the states of AZ, CA, CO, DC, FL, GA,…Official notice →
All 12 Abbott Medical recalls →

Other drugs & medical recalls in the United States

  1. 2024-11-27
  2. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number:…

    To better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors…

    FDA · DeviceOlympus Corporation of the AmericasDomestic: US Nationwide Distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Philips Medical Systems Technologies Ltd. recalls Vue PACS Diagnostic Viewer with Vue PACS client 12.1.5 to 12.2.8.454 software Model Number: 1149418 Intended use is to…

    A software issue affecting Philips Vue PACS may cause incorrect ischemic map and table value calculations when using the export function in the Perfusion Application for a Siemens CT perfusion study, resulting in…

    FDA · DevicePhilips Medical Systems Technologies Ltd.US Nationwide and Worldwide Distribution Foreign: Country Argentina Australia…Official notice →
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and…

    A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.US Nationwide and Worldwide distribution: Foreign: Algeria Algeria Argentina…Official notice →
  5. US
    Drugs & medicalClass III

    Alcon Research LLC recalls a) Alcon Custom-Pak CONSTELLATION 25G PLUS CHRISTIANA CARE HLTH SERV, # 15340-11, containing TUBING, SALEM SUMP, 14 FR; b) Alcon…

    Reported breakage in the tubing Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.

    FDA · DeviceAlcon Research LLCUS distribution to states of: Delaware, Illinois, and OklahomaOfficial notice →
  6. US
    Drugs & medicalClass II

    Roche Diagnostics Operations, Inc. recalls cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total…

    Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)

    FDA · DeviceRoche Diagnostics Operations, Inc.US: NJ, TX, NY, FL, IN, TX, OK, SC, IA,Official notice →
  7. US
    Drugs & medicalClass II

    Enchroma Inc recalls EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoor Rx,…

    Safety glasses with Rx indoor universal lenses fail to meet high impact safety specifications, which may not provide adequate protection against impact, posing a risk of eye injury.

    FDA · DeviceEnchroma IncWorldwide distribution: US (nationwide): IL, OH, NC, IA, AL, SC, NJ, TX, CO,…Official notice →