Drugs & medicalClass III

Abbott Medical recalls Merlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data from the patient s implanted…

Due to heart transmitter not being able to upgrade to the current software version in the field after experiencing a power interruption during a prior over-the-wire software download.

Read the official notice →
Date
2024-12-04
Country
United States
Brand
Abbott Medical
Product
Merlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data from the patient s implanted…
Classification
Class III
Units
Worldwide - U.S. Nationwide distribution in the states of ALABAMA, ALASKA,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abbott Medical recalls

  1. 2025-11-19
  2. US
    Drugs & medicalClass II

    Abbott Medical recalls Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400

    There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 .…

    FDA · DeviceAbbott MedicalWorldwide distribution.Official notice →
  3. 2025-04-23
  4. US
    Drugs & medicalClass II

    Abbott Medical recalls Aveir DR Leadless Pacemaker (LP) Ventricular Device REF LSP202V UDI-DI code: 05415067040725 REF LSP112V UDI-DI code: 05415067034472 The…

    Due to a manufacturing issue, leadless pacemakers may have been exposed to higher than normal temperatures which may result in premature battery failure.

    FDA · DeviceAbbott MedicalWorldwide - U.S. Nationwide distribution in the states of FL, SC, and WV and…Official notice →
  5. 2025-01-01
  6. US
    Drugs & medicalClass II

    Abbott Medical recalls The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates…

    Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the…

    FDA · DeviceAbbott MedicalWorldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →
  7. 2024-11-27
  8. US
    Drugs & medicalClass II

    Abbott Medical recalls Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile

    A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or…

    FDA · DeviceAbbott MedicalNew York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman,…Official notice →
  9. 2024-07-31
  10. US
    Drugs & medicalClass II

    Abbott Medical recalls Abbott Infinity 7, Implantable Pulse Generator, REF: 6663, SterileEO, Rx Only

    Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.

    FDA · DeviceAbbott MedicalWorldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →
  11. 2024-07-24
  12. US
    Drugs & medicalClass II

    Abbott Medical recalls Abbott Proclaim XR 5 Implantable Pulse Generator, REF 3660, Sterile EO, Rx Only

    The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.

    FDA · DeviceAbbott MedicalWorldwide - US Nationwide distribution including in the states of ALABAMA,…Official notice →
All 12 Abbott Medical recalls →

Other drugs & medical recalls in the United States

  1. 2024-12-04
  2. US
    Drugs & medicalClass III

    Remote Diagnostic Technologies Ltd. recalls Tempus LS-Manual Defibrillator Model 00-3020 The Tempus LS-Manual is a defibrillator that can deliver a shock in manual or…

    Due to incorrect labels applied to devices during servicing. Labels contain the incorrect product name and CAT number.

    FDA · DeviceRemote Diagnostic Technologies Ltd.U.S. Nationwide distribution in the states of AL, CA, FL, MO, NY, OH, TX and VA.Official notice →
  3. US
    Drugs & medicalClass II

    Shimadzu Corp. Analytical & Measuring Instruments Div. recalls Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

    There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and…

    FDA · DeviceShimadzu Corp. Analytical & Measuring Instruments Div.Worldwide - US Nationwide distribution in the states of DE, PA and the country…Official notice →
  4. US
    Drugs & medicalClass II

    Shimadzu Corp. Analytical & Measuring Instruments Div. recalls Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL

    There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and…

    FDA · DeviceShimadzu Corp. Analytical & Measuring Instruments Div.Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA,…Official notice →
  5. US
    Drugs & medicalClass II

    GE Healthcare (China) Co., Ltd. recalls GE Proteus XR/A radiographic system

    Steel cables in the Wall Stand for certain Proteus XR/a systems serviced by GE HealthCare have not been replaced at the intervals detailed in the Preventative Maintenance Action List.

    FDA · DeviceGE Healthcare (China) Co., Ltd.Worldwide - US Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    INCIPIO DEVICES recalls Offset Cup Impactor - 3/8-16 interface. Reusable surgical instrument used to position and place acetabular sup implants.

    Recalled lot does not meet specification and has reported incidents of fracture of the TIG weld.

    FDA · DeviceINCIPIO DEVICESUS Nationwide distribution in the state of TN.Official notice →
  7. US
    Drugs & medicalClass II

    Philips North America Llc recalls dS Breast 16ch 1.5T

    Potential safety issue where a patient may be harmed while preparing for or during a scan.

    FDA · DevicePhilips North America LlcUS Nationwide distribution.Official notice →
See all drugs & medical recalls →