Drugs & medicalClass II

Philips Medical Systems Technologies Ltd. recalls Vue PACS Diagnostic Viewer with Vue PACS client 12.1.5 to 12.2.8.454 software Model Number: 1149418 Intended use is to…

A software issue affecting Philips Vue PACS may cause incorrect ischemic map and table value calculations when using the export function in the Perfusion Application for a Siemens CT perfusion study, resulting in…

Read the official notice →
Date
2024-11-27
Country
United States
Brand
Philips Medical Systems Technologies Ltd.
Product
Vue PACS Diagnostic Viewer with Vue PACS client 12.1.5 to 12.2.8.454 software Model Number: 1149418 Intended use is to provide completely…
Classification
Class II
Units
US Nationwide and Worldwide Distribution Foreign: Country Argentina Australia…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

Other drugs & medical recalls in the United States

  1. 2024-11-27
  2. US
    Drugs & medicalClass II

    NuVasive Specialized Orthopedics, Inc. recalls Precice Max REF: PM10.0-80B265 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 265mm PM10.0-80B280 PRECICE MAX…

    Due to potential issues with design, manufacturing and documentation issues related to the validation of the affected products/devices.

    FDA · DeviceNuVasive Specialized Orthopedics, Inc.U.S distribution to states of: FL and NV.Official notice →
  3. US
    Drugs & medicalClass II

    Cook Incorporated recalls Approach CTO Micro Wire Guide, RPN CMW-14-190-25G, GPN G50789

    Devices from the affected device lots have labels that state the incorrect expiration dates.

    FDA · DeviceCook IncorporatedInternational Distribution to: Korea onlyOfficial notice →
  4. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number:…

    To better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors…

    FDA · DeviceOlympus Corporation of the AmericasDomestic: US Nationwide Distribution.Official notice →
  5. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and…

    A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.US Nationwide and Worldwide distribution: Foreign: Algeria Algeria Argentina…Official notice →
  6. US
    Drugs & medicalClass II

    Abbott Medical recalls Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile

    A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or…

    FDA · DeviceAbbott MedicalNew York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman,…Official notice →
  7. US
    Drugs & medicalClass III

    Alcon Research LLC recalls a) Alcon Custom-Pak CONSTELLATION 25G PLUS CHRISTIANA CARE HLTH SERV, # 15340-11, containing TUBING, SALEM SUMP, 14 FR; b) Alcon…

    Reported breakage in the tubing Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.

    FDA · DeviceAlcon Research LLCUS distribution to states of: Delaware, Illinois, and OklahomaOfficial notice →