Siemens Healthcare Diagnostics, Inc. recalls Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in…
Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH ¿2-Microglobulin (B2M) reagent.
Read the official notice →- Date
- 2024-12-11
- Country
- United States
- Brand
- Siemens Healthcare Diagnostics
- Product
- Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and…
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Argentina,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Healthcare Diagnostics recalls
- 2025-04-16
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223.
Incorrect software flagging may occur for the assay that may potentially lead to an erroneous result.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Domestic: Nationwide Distribution. International: Australia, Austria, Bahamas,…Official notice →- 2025-01-01
US
Siemens Healthcare Diagnostics, Inc. recalls Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;
Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Siemens Healthcare Diagnostics, Inc. recalls epoc NXS Host; Siemens Material Number (SMN): 11413498(JP) ; Software Version Number: 4.14.9. Intended for use by trained…
Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain…Official notice →- 2024-12-11
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica CI Analyzer. Catalog Numbers: 10947347.
Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Austria,…Official notice →- 2024-11-27
US
Siemens Healthcare Diagnostics, Inc. recalls Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and…
A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.US Nationwide and Worldwide distribution: Foreign: Algeria Algeria Argentina…Official notice →- 2024-11-20
US
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)
The IMMULITE 2000 PSA assay for use on the IMMULITE 2000 and IMMULITE 2000 XPi analyzers did not meet the High-dose Hook Effect claim as stated in the Instructions for Use (IFU).
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution.Official notice →
Other drugs & medical recalls in the United States
- 2024-12-11
US
Olympus Corporation of the Americas recalls EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190
High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO,…Official notice →US
Advanced Medical Solutions Ltd. recalls MEDLINE MAXORB EXTRA ALGINATE WOUND DRESSING 4X4 SQUARE REF MSC7044EP MEDLINE MOXORB EXTRA ALBINATE WOUND DRESSING 4X8 RECTANLE…
Due to defects on primary packaging pouch that compromise the device's ability to maintain a sterile barrier.
FDA · DeviceAdvanced Medical Solutions Ltd.U.S. Nationwide distribution in the states of IL, TX, and VA.Official notice →US
GE Healthcare (China) Co., Ltd. recalls GE Healthcare Revolution ACE, Model number 6670000-3; X-ray/computed tomography systems
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical,…
US
Cartiva, Inc recalls Cartiva Synthetic Cartilage Implant (SCI) Catalog: CAR-06-US (6mm), CAR-08-US (8mm), CAR-10-US (10mm), CAR-12-US (12mm)
Patients implanted with synthetic cartilage implant, may experience a higher-than expected occurrence rate of the following hazards: revision, removal, implant subsidence, displacement, pain, nerve damage or…
FDA · DeviceCartiva, IncWorldwide - US Nationwide distribution in the states of CO, CT, FL, IL, MI,…Official notice →US
Siemens Healthcare Diagnostics, Inc. recalls Atellica CI Analyzer. Catalog Numbers: 10947347.
Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Austria,…Official notice →US
Becton Dickinson & Co. recalls BD COR GX Instrument Version or Model: 443990 Catalog Number: 443990
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is…
FDA · DeviceBecton Dickinson & Co.Worldwide distribution - US Nationwide and the countries of Algeria,…Official notice →