Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)
The IMMULITE 2000 PSA assay for use on the IMMULITE 2000 and IMMULITE 2000 XPi analyzers did not meet the High-dose Hook Effect claim as stated in the Instructions for Use (IFU).
Read the official notice →- Date
- 2024-11-20
- Country
- United States
- Brand
- Siemens Healthcare Diagnostics
- Product
- IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Healthcare Diagnostics recalls
- 2024-12-11
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in…
Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH ¿2-Microglobulin (B2M) reagent.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →US
Siemens Healthcare Diagnostics, Inc. recalls Atellica CI Analyzer. Catalog Numbers: 10947347.
Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Austria,…Official notice →- 2024-11-27
US
Siemens Healthcare Diagnostics, Inc. recalls Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and…
A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.US Nationwide and Worldwide distribution: Foreign: Algeria Algeria Argentina…Official notice →- 2024-11-13
US
Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from…
- 2024-10-09
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM aTPO Assay, Material Number 10995466 (100 Test) and 10995467 (500 Test)
Positive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide.Official notice →- 2024-09-25
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Microalbumin_2 (¿ALB_2)-For in vitro diagnostic use in the quantitative measurement of microalbumin in human…
Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH,…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
Other drugs & medical recalls in the United States
- 2024-11-27
US
Padagis US LLC recalls Triamcinolone Acetonide Cream USP, 0.025%, 1 LB (454 g) per jar, Rx Only, Manufactured By Padagis, Minneapolis, MN 55427. NDC: 45802-0063-05
Subpotent and Superpotent Drug. Out of specification assay results recorded as part of Uniformity of Container test during long-term stability testing.
US
- 2024-11-20
US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline Iris Scissors, Single-Use, Curved, Standard 4.5 , Model Nos. DYNJ04049 (cases), DYNJ04049H (eaches)
Tip protector may fall off within the packaging, resulting in compromised sterile barrier system and loss of sterility.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distributions in the states of AL, AR, CO, CT, FL, IA, IN, KS,…Official notice →US
Getinge Usa Sales Inc recalls MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical…
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may…
FDA · DeviceGetinge Usa Sales IncWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
AVID Medical, Inc. recalls Halyard KIT, SHOULDER - Medical convenience kits Model Number: ESJH014-06
Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or…
US
Medtronic Neuromodulation recalls Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable…
The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.
FDA · DeviceMedtronic NeuromodulationWorldwide - US Nationwide distribution in the states of AL, AZ, CA, FL, GA,…Official notice →