PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842,…
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages…
Read the official notice →- Date
- 2025-10-08
- Country
- United States
- Brand
- Philips Medical Systems Nederland
- Product
- Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979,…
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips Medical Systems Nederland recalls
- 2026-01-21
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10F; Model Number: 722002;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →- 2025-12-03
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Domestic: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MD, MN, MO, NM, NY,…Official notice →- 2025-10-29
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system
Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Distribution US states of MO and TX; Belgium, Canada, Fiji, Germany, India,…Official notice →- 2025-10-01
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638,…
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Albania, Algeria,…Official notice →- 2025-09-17
US
Philips Medical Systems Nederland B.V. recalls IntelliSpace Cardiovascular, Software 8.0.0.4.
Software issue that results in the display of outdated information.
FDA · DevicePhilips Medical Systems Nederland B.V.US Nationwide distribution in the states of GA, NC & TX.Official notice →- 2025-09-10
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Interventional Fluoroscopic X-Ray System - Philips Allura Xper Series. Labeled as the following with corresponding model…
Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide distribution in the states of AL, AZ, CA, DC, FL,…Official notice →
Other drugs & medical recalls in the United States
- 2025-10-08
US
Northeast Scientific Inc. recalls NES Reprocessed 1.7mm RX Turbo Elite Laser Atherectomy Catheter. Model Number: R-417-156. The NES Reprocessed Turbo-Elite Laser…
Potential for breaches in the sterile barrier packaging, compromising sterility assurance.
FDA · DeviceNortheast Scientific Inc.US Nationwide distribution to CA, FL, IL, KS, LA, MD, MI, MO, NC, NJ, NM, NV,…Official notice →US
S.I.N. Implant System Ltda recalls IMPLANT EPIKUT PLUS MT16 3,5X10,0MM, REF: ILM 3510N; IMPLANT EPIKUT PLUS MT16 3.8X11.5MM, REF: ILM 3511N; IMPLANT EPIKUT PLUS MT16…
Dental implants were imported under temperature and storage conditions that do not comply with the instructions for use and product labeling so there may be packaging damage, which could compromise product quality and…
FDA · DeviceS.I.N. Implant System LtdaUS Nationwide distribution including in the states of TX, IL, KY, NC, TN, CO,…Official notice →US
Getinge Disinfection Ab recalls Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
FDA · DeviceGetinge Disinfection AbWorldwide - US Nationwide distribution in the states of Colorado,…Official notice →US
Apotex Corp. recalls Brimonidine Tartrate/Timolol Maleate Ophthalmic Solution, 0.2%/0.5%, a) 5 mL and b) 10 mL, Rx Only, Sterile, Manufactured by: Apotex Inc.,…
Lack of Assurance of Sterility; atypical weight loss due to improper bottle sealing leading to potential sterility concerns
US
US
Zydus Pharmaceuticals (USA) Inc recalls Entecavir Tablets, USP, 0.5 mg, 30-count bottle, Rx Only, Manufactured by: Zydus Lifesciences Ltd., Ahmedabad, India,…
Failed impurity/degradation specifications:Out of Specification result for an individual organic impurity