Drugs & medicalClass II

Northeast Scientific Inc. recalls NES Reprocessed 1.7mm RX Turbo Elite Laser Atherectomy Catheter. Model Number: R-417-156. The NES Reprocessed Turbo-Elite Laser…

Potential for breaches in the sterile barrier packaging, compromising sterility assurance.

Read the official notice →
Date
2025-10-08
Country
United States
Brand
Northeast Scientific Inc.
Product
NES Reprocessed 1.7mm RX Turbo Elite Laser Atherectomy Catheter. Model Number: R-417-156. The NES Reprocessed Turbo-Elite Laser…
Classification
Class II
Units
US Nationwide distribution to CA, FL, IL, KS, LA, MD, MI, MO, NC, NJ, NM, NV,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

Other drugs & medical recalls in the United States

  1. 2025-10-08
  2. US
    Drugs & medicalClass II

    Beckman Coulter Inc. recalls DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i…

    On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress…

    FDA · DeviceBeckman Coulter Inc.Internationally distribution to the countries of Poland, India, Dominican…Official notice →
  3. US
    Drugs & medicalClass II

    CHANGE HEALTHCARE CANADA COMPANY recalls Change Healthcare Radiology Solutions software version 14.2.2

    Due to software issue, radiology reports may not be fully displayed when viewing.

    FDA · DeviceCHANGE HEALTHCARE CANADA COMPANYWorldwide - US Nationwide distribution in the states of AR, WI, NJ, IN, AL,…Official notice →
  4. US
    Drugs & medicalClass II

    Aesculap AG recalls AESCULAP ELAN 4 FIXED DURAGUARD STANDARD. Model Number: GB942R.

    Mislabeling. The ELAN 4 FIXED DURAGUARD STANDARD was labeled as "Long" and the ELAN 4 FIXED DURAGUARD LONG was labeled as "Standard."

    FDA · DeviceAesculap AGUS distribution to Missouri. OUS International distribution to Norway,…Official notice →
  5. US
    Drugs & medicalClass II

    S.I.N. Implant System Ltda recalls IMPLANT EPIKUT PLUS MT16 3,5X10,0MM, REF: ILM 3510N; IMPLANT EPIKUT PLUS MT16 3.8X11.5MM, REF: ILM 3511N; IMPLANT EPIKUT PLUS MT16…

    Dental implants were imported under temperature and storage conditions that do not comply with the instructions for use and product labeling so there may be packaging damage, which could compromise product quality and…

    FDA · DeviceS.I.N. Implant System LtdaUS Nationwide distribution including in the states of TX, IL, KY, NC, TN, CO,…Official notice →
  6. US
    Drugs & medicalClass II

    Getinge Disinfection Ab recalls Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

    Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

    FDA · DeviceGetinge Disinfection AbWorldwide - US Nationwide distribution in the states of Colorado,…Official notice →
  7. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842,…

    The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages…

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Albania, Algeria,…Official notice →