PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied…
Read the official notice →- Date
- 2025-12-03
- Country
- United States
- Brand
- Philips Medical Systems Nederland
- Product
- Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
- Classification
- Class II
- Units
- Domestic: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MD, MN, MO, NM, NY,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips Medical Systems Nederland recalls
- 2026-01-21
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and…
Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Afghanistan,…Official notice →US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model…
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Albania, Algeria,…Official notice →US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10F; Model Number: 722002;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →- 2025-10-29
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system
Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Distribution US states of MO and TX; Belgium, Canada, Fiji, Germany, India,…Official notice →- 2025-10-08
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842,…
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Albania, Algeria,…Official notice →- 2025-10-01
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638,…
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Albania, Algeria,…Official notice →
Other drugs & medical recalls in the United States
- 2025-12-03
US
INTELERAD MEDICAL SYSTEMS INCORPORATED recalls IntelePACS - InteleConnect / TechPortal
Software application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant Clinical Info field when modifying…
FDA · DeviceINTELERAD MEDICAL SYSTEMS INCORPORATEDUS Nationwide distribution in the states of AZ, CT, AL, IL, WA, MD, NJ,CA, TN,…Official notice →US
DJO, LLC recalls Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02
The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures…
FDA · DeviceDJO, LLCUS Nationwide distribution in the states of WI, TN, TX, MD, GA, UT, WA, MI,…Official notice →US
Mindray DS USA, Inc. dba Mindray North America recalls Sevoflurane (SEVO) V90 Electronic Vaporizer with the A9 Anesthesia System. Labeled as the following: 1. Sev…
Potential for anesthesia leakage.
FDA · DeviceMindray DS USA, Inc. dba Mindray North AmericaWorldwide - US Nationwide distribution in the states of California, Florida,…Official notice →US
B Braun Medical Inc recalls Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA · DeviceB Braun Medical IncWorldwide distribution - US distribution and international distribution to…Official notice →US
BioFire Diagnostics, LLC recalls BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit)
Due to unintended movement of pouch chemistry, nucleic acid test may result in elevated risk of false negative results and risk of control failures.
FDA · DeviceBioFire Diagnostics, LLCUS:AZ, FL,GA,ID NC, NH, NJ, NY, OR, PA, SC OUS: NoneOfficial notice →US
Stryker Corporation recalls NICO BrainPath; Product Number/Product Name NN-8024 BrainPath Shephard's Hook - Greenberg; NN-8025 BrainPath Shephard's Hook - Sugita;…
Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.