PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638,…
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system…
Read the official notice →- Date
- 2025-10-01
- Country
- United States
- Brand
- Philips Medical Systems Nederland
- Product
- Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979,…
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips Medical Systems Nederland recalls
- 2025-12-03
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Domestic: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MD, MN, MO, NM, NY,…Official notice →- 2025-10-29
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system
Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Distribution US states of MO and TX; Belgium, Canada, Fiji, Germany, India,…Official notice →- 2025-10-08
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842,…
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Albania, Algeria,…Official notice →- 2025-09-17
US
Philips Medical Systems Nederland B.V. recalls IntelliSpace Cardiovascular, Software 8.0.0.4.
Software issue that results in the display of outdated information.
FDA · DevicePhilips Medical Systems Nederland B.V.US Nationwide distribution in the states of GA, NC & TX.Official notice →- 2025-09-10
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Interventional Fluoroscopic X-Ray System - Philips Allura Xper Series. Labeled as the following with corresponding model…
Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide distribution in the states of AL, AZ, CA, DC, FL,…Official notice →US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion System configured with a 1 Phase UPS. Azurion family (R1.x, R2.x). Labeled as the following with…
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
Other drugs & medical recalls in the United States
- 2025-10-08
US
US
Zydus Pharmaceuticals (USA) Inc recalls Entecavir Tablets, USP, 0.5 mg, 30-count bottle, Rx Only, Manufactured by: Zydus Lifesciences Ltd., Ahmedabad, India,…
Failed impurity/degradation specifications:Out of Specification result for an individual organic impurity
- 2025-10-01
US
Maquet Cardiovascular, LLC recalls Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH)…
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft,…
FDA · DeviceMaquet Cardiovascular, LLCUS Nationwide distribution in the states of FL, IL, MD & TX.Official notice →US
Mindray DS USA, Inc. dba Mindray North America recalls BeneVision N1 Patient Monitor (N1). Part Numbers: 6660E-PA00017, 6660E-PA00014, 6660E-PA00012, 6660E-PA00011,…
Potential for activation of an abnormal alarm pause.
FDA · DeviceMindray DS USA, Inc. dba Mindray North AmericaWorldwide - Nationwide distribution and the country of Canada.Official notice →US
Tandem Diabetes Care, Inc. recalls t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)
an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent…
FDA · DeviceTandem Diabetes Care, Inc.International distribution to the country of Israel.Official notice →US
Stryker Communications recalls Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034
Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur.
FDA · DeviceStryker CommunicationsUS Nationwide distribution in the states of TX, IL, NY, UT, OR, AZ, VT, NY,…Official notice →