Drugs & medicalClass II

Maquet Cardiovascular, LLC recalls Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH)…

Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft,…

Read the official notice →
Date
2025-10-01
Country
United States
Brand
Maquet Cardiovascular
Product
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH)…
Classification
Class II
Units
US Nationwide distribution in the states of FL, IL, MD & TX.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Maquet Cardiovascular recalls

  1. 2026-07-22
  2. US
    Drugs & medicalClass II

    Maquet Cardiovascular, LLC recalls The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II…

    Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its…

    FDA · DeviceMaquet Cardiovascular, LLCUS Nationwide distribution in the states of NY, CA, SC, FL, TX, GA, OR.Official notice →
  3. 2025-09-24
  4. US
    Drugs & medicalClass II

    Maquet Cardiovascular, LLC recalls Heartstring III Proximal Seal System. Intravascular anastomosis occluder.

    Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate…

    FDA · DeviceMaquet Cardiovascular, LLCWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  5. 2025-04-30
  6. US
    Drugs & medicalClass II

    Maquet Cardiovascular, LLC recalls The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm extended length…

    There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while…

    FDA · DeviceMaquet Cardiovascular, LLCWorldwide distribution - US Nationwide and the countries of Mexico, United…Official notice →
  7. 2025-04-23
  8. US
    Drugs & medicalClass II

    Maquet Cardiovascular, LLC recalls Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).

    Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.

    FDA · DeviceMaquet Cardiovascular, LLCWorldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL,…Official notice →
  9. 2025-01-15
  10. US
    Drugs & medicalClass I

    Maquet Cardiovascular, LLC recalls VasoView HemoPro 2 Endoscopic Vessel Harvesting System. Model Numbers: VH-4000. Indicated for use in minimally invasive surgery…

    Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool

    FDA · DeviceMaquet Cardiovascular, LLCNationwide distribution. International distribution to Australia, Austria,…Official notice →
  11. 2024-11-27
  12. US
    Drugs & medicalClass II

    Maquet Cardiovascular, LLC recalls Acrobat-i Vacuum Stabilizer System. Model Number C-OM-10000Z

    Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.

    FDA · DeviceMaquet Cardiovascular, LLCNationwide distribution. International distribution to Albania, Algeria,…Official notice →
All 23 Maquet Cardiovascular recalls →

Other drugs & medical recalls in the United States

  1. 2025-10-08
  2. US
    Drugs & medicalClass II

    Apotex Corp. recalls Brimonidine Tartrate/Timolol Maleate Ophthalmic Solution, 0.2%/0.5%, a) 5 mL and b) 10 mL, Rx Only, Sterile, Manufactured by: Apotex Inc.,…

    Lack of Assurance of Sterility; atypical weight loss due to improper bottle sealing leading to potential sterility concerns

    FDA · DrugApotex Corp.Nationwide in the USOfficial notice →
  3. US
  4. US
    Drugs & medicalClass II

    Zydus Pharmaceuticals (USA) Inc recalls Entecavir Tablets, USP, 0.5 mg, 30-count bottle, Rx Only, Manufactured by: Zydus Lifesciences Ltd., Ahmedabad, India,…

    Failed impurity/degradation specifications:Out of Specification result for an individual organic impurity

    FDA · DrugZydus Pharmaceuticals (USA) IncDistributed Nationwide in the USAOfficial notice →
  5. 2025-10-01
  6. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638,…

    Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system…

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Albania, Algeria,…Official notice →
  7. US
    Drugs & medicalClass II

    Mindray DS USA, Inc. dba Mindray North America recalls BeneVision N1 Patient Monitor (N1). Part Numbers: 6660E-PA00017, 6660E-PA00014, 6660E-PA00012, 6660E-PA00011,…

    Potential for activation of an abnormal alarm pause.

    FDA · DeviceMindray DS USA, Inc. dba Mindray North AmericaWorldwide - Nationwide distribution and the country of Canada.Official notice →
  8. US
    Drugs & medicalClass II

    Tandem Diabetes Care, Inc. recalls t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)

    an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent…

    FDA · DeviceTandem Diabetes Care, Inc.International distribution to the country of Israel.Official notice →