Drugs & medicalClass II

Maquet Cardiovascular, LLC recalls The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II…

Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its…

Read the official notice →
Date
2026-07-22
Country
United States
Brand
Maquet Cardiovascular
Product
The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device…
Classification
Class II
Units
US Nationwide distribution in the states of NY, CA, SC, FL, TX, GA, OR.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Maquet Cardiovascular recalls

  1. 2025-10-01
  2. US
    Drugs & medicalClass II

    Maquet Cardiovascular, LLC recalls Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH)…

    Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft,…

    FDA · DeviceMaquet Cardiovascular, LLCUS Nationwide distribution in the states of FL, IL, MD & TX.Official notice →
  3. 2025-09-24
  4. US
    Drugs & medicalClass II

    Maquet Cardiovascular, LLC recalls Heartstring III Proximal Seal System. Intravascular anastomosis occluder.

    Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate…

    FDA · DeviceMaquet Cardiovascular, LLCWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  5. 2025-04-30
  6. US
    Drugs & medicalClass II

    Maquet Cardiovascular, LLC recalls The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm extended length…

    There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while…

    FDA · DeviceMaquet Cardiovascular, LLCWorldwide distribution - US Nationwide and the countries of Mexico, United…Official notice →
  7. 2025-04-23
  8. US
    Drugs & medicalClass II

    Maquet Cardiovascular, LLC recalls Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).

    Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.

    FDA · DeviceMaquet Cardiovascular, LLCWorldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL,…Official notice →
  9. 2025-01-15
  10. US
    Drugs & medicalClass I

    Maquet Cardiovascular, LLC recalls VasoView HemoPro 2 Endoscopic Vessel Harvesting System. Model Numbers: VH-4000. Indicated for use in minimally invasive surgery…

    Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool

    FDA · DeviceMaquet Cardiovascular, LLCNationwide distribution. International distribution to Australia, Austria,…Official notice →
  11. 2024-11-27
  12. US
    Drugs & medicalClass II

    Maquet Cardiovascular, LLC recalls Acrobat-i Vacuum Stabilizer System. Model Number C-OM-10000Z

    Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.

    FDA · DeviceMaquet Cardiovascular, LLCNationwide distribution. International distribution to Albania, Algeria,…Official notice →
All 23 Maquet Cardiovascular recalls →

Other drugs & medical recalls in the United States

  1. 2026-07-29
  2. US
  3. 2026-07-22
  4. US
    Drugs & medicalClass II

    Tosoh Bioscience Inc recalls ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh…

    The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone…

    FDA · DeviceTosoh Bioscience IncGlobal Distribution. US Nationwide.Official notice →
  5. US
    Drugs & medicalClass II

    Daavlin Distributing Company recalls Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

    Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

    FDA · DeviceDaavlin Distributing CompanyUS Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    3B Medical, Inc. recalls Brand Name: Rio" II Product Name: Rio" II Nasal Pillows Interface Model/Catalog Number: RII1001 (Small), RII1002 (Medium), RII1003 (Large)…

    Packaging and labeling discrepancy involving product size identification and associated Unique Device Identifier (UDI/GTIN).

    FDA · Device3B Medical, Inc.US Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Medtronic Perfusion Systems recalls Medtronic Bio-Medicus Adult Cannula Kit, Product Number/CFN 96530-115; Cardiopulmonary bypass vascular cannula

    Medtronic received two complaints of loose foreign material confirmed as containing trace levels of blood.

    FDA · DeviceMedtronic Perfusion SystemsWorldwide - US Nationwide distribution in the states of MD, CA, MI, WA, PA,…Official notice →
  8. US
    Drugs & medicalClass II

    Medline Industries, LP recalls Medline Kits containing Zimmer Biomet mixing bowls and spatulas TOTAL KNEE PACK DYNJT6267

    Kits contain Zimmer Biomet mixing bowls and spatulas, which were recalled due to seal nonconformities which present risk of sterility breach. Health consequences from use of product include clinically insignificant…

    FDA · DeviceMedline Industries, LPUS Nationwide distribution in the states of CA and UT.Official notice →
See all drugs & medical recalls →