Hamilton Medical AG recalls HAMILTON-H900 Humidifier, Models: 950001, 950004
When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm…
Read the official notice →- Date
- 2022-07-20
- Country
- United States
- Brand
- Hamilton Medical
- Product
- HAMILTON-H900 Humidifier, Models: 950001, 950004
- Classification
- Class III
- Units
- US: WI, FL, CA, NV, PA, TX, PA
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Hamilton Medical recalls
- 2024-06-26
US
Hamilton Medical AG recalls Ventilator HAMILTON-C6, PN: 160021
Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube…
FDA · DeviceHamilton Medical AGUS Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR,…Official notice →- 2023-09-27
US
Hamilton Medical, Inc. recalls Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption…
FDA · DeviceHamilton Medical, Inc.Distribution US nationwide, including Puerto Rico. OUS: Not providedOfficial notice →- 2023-08-02
US
Hamilton Medical, Inc. recalls HAMILTON C3 Ventilator REF 160005 Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally…
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
FDA · DeviceHamilton Medical, Inc.U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware…Official notice →- 2023-05-03
US
Hamilton Medical AG recalls HAMILTON-C6, REF: 160021
Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an…
FDA · DeviceHamilton Medical AGUS Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA,…Official notice →- 2022-11-30
US
Hamilton Medical AG recalls Hamilton-C6
Due to a malfunction related to the backlight on the ventilator screen. The screen goes black for approximately 2-3 seconds, but it brightens up again when the backlight switches on again.
FDA · DeviceHamilton Medical AGU.S. Nationwide distribution including in the states of CA, FL, GA, ID, IN,…Official notice →- 2022-08-31
US
Hamilton Medical AG recalls HAMILTON-C6 Intensive Care Ventilator, REF: 160021
Ventilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety…
Other drugs & medical recalls in the United States
- 2022-07-27
US
Glenmark Pharmaceuticals Inc., USA recalls Aspirin and Extended-Release Dipyridamole Capsules, 25mg/200 mg, 60-count bottle, Rx Only, Manufactured for: Glenmark…
Failed Tablet/Capsule Specification : Capsule breakage
US
US
Vi-Jon, LLC recalls Hydrogen Peroxide Topical Solution USP, 3%, 16 FL OZ (1 PT) 473 mL bottle, packaged as a) Signature Care NDC 21130-871-43 UPC 3…
Cross Contamination With Other Products: Product is cross contaminated with low levels of isopropyl alcohol.
- 2022-07-20
US
Agilent Technologies, Inc. recalls Kit Label: Anti-FITC-AP CISH Accessory Kit (Dako Omnis), Box A REF K589911-2 Anti-FITC-AP CISH Accessory Kit (Dako Omnis), Box A REF…
Label provided for substrate vail contained incorrect expiration date.
FDA · DeviceAgilent Technologies, Inc.U.S. Nationwide distribution in the states of NY and VA.Official notice →US
Aesculap Implant Systems LLC recalls AESCULAP TC COOELY MICRO NEEDLEHLDER SERR 200MM, product code MG524R. manual surgical instrument designed to grasp a suture needle
Needleholder incorrectly labeled as 200MM when the actual instrument size is 150MM.
US
Breas Medical, Inc. recalls Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require…
A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health consequences could…