Drugs & medicalClass I

Hamilton Medical, Inc. recalls HAMILTON C3 Ventilator REF 160005 Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally…

Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.

Read the official notice →
Date
2023-08-02
Country
United States
Brand
Hamilton Medical
Product
HAMILTON C3 Ventilator REF 160005 Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally…
Classification
Class I
Units
U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Hamilton Medical recalls

  1. 2025-08-06
  2. US
    Drugs & medicalClass I

    Hamilton Medical AG recalls Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and…

    Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete…

    FDA · DeviceHamilton Medical AGUS Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC,…Official notice →
  3. 2024-06-26
  4. US
    Drugs & medicalClass I

    Hamilton Medical AG recalls Ventilator HAMILTON-C6, PN: 160021

    Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube…

    FDA · DeviceHamilton Medical AGUS Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR,…Official notice →
  5. 2023-09-27
  6. US
    Drugs & medicalClass I

    Hamilton Medical, Inc. recalls Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator

    Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption…

    FDA · DeviceHamilton Medical, Inc.Distribution US nationwide, including Puerto Rico. OUS: Not providedOfficial notice →
  7. 2023-05-03
  8. US
    Drugs & medicalClass II

    Hamilton Medical AG recalls HAMILTON-C6, REF: 160021

    Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an…

    FDA · DeviceHamilton Medical AGUS Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA,…Official notice →
  9. 2022-11-30
  10. US
    Drugs & medicalClass II

    Hamilton Medical AG recalls Hamilton-C6

    Due to a malfunction related to the backlight on the ventilator screen. The screen goes black for approximately 2-3 seconds, but it brightens up again when the backlight switches on again.

    FDA · DeviceHamilton Medical AGU.S. Nationwide distribution including in the states of CA, FL, GA, ID, IN,…Official notice →
  11. 2022-08-31
  12. US
    Drugs & medicalClass I

    Hamilton Medical AG recalls HAMILTON-C6 Intensive Care Ventilator, REF: 160021

    Ventilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety…

    FDA · DeviceHamilton Medical AGUS nationwide distribution including Puerto Rico.Official notice →
All 11 Hamilton Medical recalls →

Other drugs & medical recalls in the United States

  1. 2023-08-02
  2. US
    Drugs & medicalClass III

    Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products NT-proBNP II Range Verifiers-For in vitro diagnostic use only. For use in verifying the…

    May not meet the stability specifications for shelf life outlined in the Instructions For Use (IFU).

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.US Nationwide distribution in the state of CA.Official notice →
  3. US
    Drugs & medicalClass II

    Datascope Corp. recalls LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product…

    During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending…

    FDA · DeviceDatascope Corp.Domestic distribution nationwide. International distribution worldwide.Official notice →
  4. US
    Drugs & medicalClass II

    Qiagen GmbH recalls QIAstat-Dx¿ Respiratory SARS-CoV-2 Panel -IVD intended for the detection and differentiation of nucleic acid from SARS-CoV-2 and the…

    If cartridges of the affected Serial numbers (SNs) are used, in most cases, the error codes 0xY0003EA and 0xY00045A (where Y is a number between 1 and 4) that lead to a run abortion will occur and no results will be…

    FDA · DeviceQiagen GmbHAL, AR, CA, DC, FL, GA, ID, KS, KY, MD, MN, NJ, NY,PA, SC, TX, VAOfficial notice →
  5. US
  6. US
    Drugs & medicalClass II

    New Vitalis Pharmacy LLC dba New Vitalis Pharmacy recalls Testosterone Cypionate-Testosterone Propionate 180-20 mg/mL in cottonseed oil, Compounded Rx, For…

    Lack of Assurance of Sterility

    FDA · DrugNew Vitalis Pharmacy LLC dba New Vitalis PharmacyNationwideOfficial notice →
  7. US
    Drugs & medicalClass II

    Kramer Laboratories, Inc. recalls Safe Tussin DM Cough + Chest Congestion (Dextromethorphan HBr/ Cough Suppressant Guaifenesin/Expectorant), packaged in 4.0 FL…

    CGMP Deviations: use of non-food grade lubricant in mixing vessel.

    FDA · DrugKramer Laboratories, Inc.nationwide within the United StatesOfficial notice →
See all drugs & medical recalls →