Hamilton Medical AG recalls Hamilton-C6
Due to a malfunction related to the backlight on the ventilator screen. The screen goes black for approximately 2-3 seconds, but it brightens up again when the backlight switches on again.
Read the official notice →- Date
- 2022-11-30
- Country
- United States
- Brand
- Hamilton Medical
- Product
- Hamilton-C6
- Classification
- Class II
- Units
- U.S. Nationwide distribution including in the states of CA, FL, GA, ID, IN,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Hamilton Medical recalls
- 2024-06-26
US
Hamilton Medical AG recalls Ventilator HAMILTON-C6, PN: 160021
Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube…
FDA · DeviceHamilton Medical AGUS Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR,…Official notice →- 2023-09-27
US
Hamilton Medical, Inc. recalls Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption…
FDA · DeviceHamilton Medical, Inc.Distribution US nationwide, including Puerto Rico. OUS: Not providedOfficial notice →- 2023-08-02
US
Hamilton Medical, Inc. recalls HAMILTON C3 Ventilator REF 160005 Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally…
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
FDA · DeviceHamilton Medical, Inc.U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware…Official notice →- 2023-05-03
US
Hamilton Medical AG recalls HAMILTON-C6, REF: 160021
Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an…
FDA · DeviceHamilton Medical AGUS Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA,…Official notice →- 2022-08-31
US
Hamilton Medical AG recalls HAMILTON-C6 Intensive Care Ventilator, REF: 160021
Ventilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety…
- 2022-07-20
US
Hamilton Medical AG recalls HAMILTON-H900 Humidifier, Models: 950001, 950004
When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm…
Other drugs & medical recalls in the United States
- 2022-11-30
US
Endo-Therapeutics, Inc. recalls eSuction Small Cavity, Model ET2005
Improper device regulatory classification
US
Beckman Coulter, Inc. recalls MicroScan autoSCAN-4 Instrument, Catalog Number B1018-280
Due to MicroScan autoSCAN-4 Instruments being manufactured with an incorrect diffuser plate.
FDA · DeviceBeckman Coulter, Inc.Worldwide - U.S. Nationwide distribution in the states of AK, FL, IA, ID, MI…Official notice →US
Abbott Molecular, Inc. recalls Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use
Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flapper was found to not open when the Systems Solution drawer was closed and…
US
Allison Medical, Inc. recalls InControl Insulin Syringes, Short Needle, 29 Gauge, 1CC
Insulin syringe retail packages incorrectly labeled as "Short Needle".
US
Medical Action Industries, Inc. 306 recalls B. Braun CARESITE Port Access Kit w/Tegaderm, REF 375214, Central Venous (cardiac) Catheter Tray (REF number corrected…
The kits were incorrectly labeled with an extended expiration date of 02/21/2024 instead of 02/10/2024.
FDA · DeviceMedical Action Industries, Inc. 306US Nationwide distribution in the state of Pennsylvania.Official notice →US
3M Healthcare Business recalls 3M Steri-Drape, Small Drape with Adhesive Aperture, REF 1020, general surgery drape
During a recent investigation, 3M confirmed the liner on the adhesive component of the affected drapes is difficult to remove without damaging the product and may render the product unusable. Additionally, 3M has…