Aesculap Implant Systems LLC recalls AESCULAP TC COOELY MICRO NEEDLEHLDER SERR 200MM, product code MG524R. manual surgical instrument designed to grasp a suture needle
Needleholder incorrectly labeled as 200MM when the actual instrument size is 150MM.
Read the official notice →- Date
- 2022-07-20
- Country
- United States
- Brand
- Aesculap Implant Systems
- Product
- AESCULAP TC COOELY MICRO NEEDLEHLDER SERR 200MM, product code MG524R. manual surgical instrument designed to grasp a suture needle
- Classification
- Class II
- Units
- United States: TX and AK
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Aesculap Implant Systems recalls
- 2022-07-27
US
Aesculap Implant Systems LLC recalls ENNOVATE MIS REMOVALKEY SHORT, product code SZ380R
Fracture or breakage of the spinal fixation arm at the downtube instrument if the 90 degree angle is not met when using the removal key with the spinal fixation system.
FDA · DeviceAesculap Implant Systems LLCUnited States Nationwide distribution in the states of CA, CO, IL, MI, OH & TX.Official notice →US
Aesculap Implant Systems LLC recalls MODULIFT VBR SZ.M 4MM SET SCREW DRIVER, MF790R, part of Modulift VBR Medium Instrument Set, ST-0441
Set screwdriver fails during surgery by deforming at the tip during engagement and manipulation of the implant.
US
Aesculap Implant Systems LLC recalls Universal Robot UR5 arm, PV010204 subcomponent of Aesculap Aeos Digital Microscope System, PV010
Malfunction of the robotic arm preventing further movement of the robotic arm and requiring restart of the robotic digital microscope system.
US
Aesculap Implant Systems LLC recalls RIGHT CORONAL BENDER, TEK1940. Instrument used during spinal surgery.
Customized coronal rod benders may deform the implant when used with the spinal system devices.
US
Aesculap Implant Systems LLC recalls ENDURO SPECIAL KEY TIB.LOCK.RING F1/10MM product code NP462R used with EnduRo Total Knee System. instrument used during orthopedic…
Nonfunctional key due to the potential for damage to the key pins cause by force of use during assembly of the tibia locking ring.
FDA · DeviceAesculap Implant Systems LLCUS Nationwide distribution in the states of AL, AZ, CA, CO, ID, IL, LA, MA,…Official notice →- 2021-10-06
US
Aesculap Implant Systems LLC recalls PAS-PORT PROXIMAL ANASTOMOSIS SYSTEM Catalog Number: FC700SU
Malfuncton-Implant did not deploy successfully may necessitate the need for an additional medical intervention for patients which would result in a potential delay in surgery
FDA · DeviceAesculap Implant Systems LLCUS Nationwide distribution in the states of FL, MI, NY, TX, WI.Official notice →
Other drugs & medical recalls in the United States
- 2022-07-27
US
Vi-Jon, LLC recalls Hydrogen Peroxide Topical Solution USP, 3%, 16 FL OZ (1 PT) 473 mL bottle, packaged as a) Signature Care NDC 21130-871-43 UPC 3…
Cross Contamination With Other Products: Product is cross contaminated with low levels of isopropyl alcohol.
- 2022-07-20
US
Hamilton Medical AG recalls HAMILTON-H900 Humidifier, Models: 950001, 950004
When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm…
US
Agilent Technologies, Inc. recalls Kit Label: Anti-FITC-AP CISH Accessory Kit (Dako Omnis), Box A REF K589911-2 Anti-FITC-AP CISH Accessory Kit (Dako Omnis), Box A REF…
Label provided for substrate vail contained incorrect expiration date.
FDA · DeviceAgilent Technologies, Inc.U.S. Nationwide distribution in the states of NY and VA.Official notice →US
Breas Medical, Inc. recalls Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require…
A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health consequences could…
US
Ivenix, Inc. recalls Ivenix Infusion System (IIS)
Downstream occlusion alarm is sometimes immediately followed by a pump problem alarm.
US
Waldemar Link GmbH & Co. KG (Mfg Site) recalls Tibial Component Modular. orthopedic prosthesis. Model Nos: Small, W 16-2817/32 Medium, W 16-2817/35 Large, W 16-2817/37
There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.
FDA · DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Domestic distribution to AL, CA, FL, GA, IL, IN, KS, MI, MN, NV, NY, OH, TN,…Official notice →