Drugs & medicalClass II

Fresenius Medical Care Holdings, Inc. recalls Novalung Sensor Box (F30000163) a component of Novalung Console (F30000162)- Indicated for long-term (> 6 hours)…

Error messages 206 (yellow) and 208 (red) technical failure, flow measurement during use of the Novalung System imply that the communication between flow sensor and sensor box is interrupted. An interrupted…

Read the official notice →
Date
2022-02-16
Country
United States
Brand
Fresenius Medical Care Holdings
Product
Novalung Sensor Box (F30000163) a component of Novalung Console (F30000162)- Indicated for long-term (> 6 hours) respiratory/…
Classification
Class II
Units
US Nationwide distribution in the states of AL, AZ, CA, CO, FL, IL, MI, NJ,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Fresenius Medical Care Holdings recalls

  1. 2022-09-28
  2. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius…

    Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV,…Official notice →
  3. 2022-08-03
  4. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls AquaC UNO H [120V, 60Hz, 1700VA, Fresenius US-Version: 24-0970-1 Vivonic Part Number: G02000966-US

    Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on…

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution.Official notice →
  5. 2022-07-20
  6. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls NATURALYTE Liquid Acid 3.0K 3.0Ca- indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for…

    Alternate Bottle Caps labeled with "Sealed for your Protection" on the foil not qualified for use

    FDA · DeviceFresenius Medical Care Holdings, Inc.NatiowideOfficial notice →
  7. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ…

    AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device.…

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of NC and WV.Official notice →
  8. 2022-07-13
  9. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls NaturaLyte 2301 2K 3.0 Ca 1 Mg Gal, Model No. 08-2301-3

    The product was potentially exposed to below-recommended storage temperatures, which may cause leaks in the packaging. Leaks can lead to biological contamination and therefore an increased risk of potential bloodstream…

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of Texas and Louisiana.Official notice →
  10. 2022-07-06
  11. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls Optiflux 200NRe Dialyzer Finished Assy., Model no. 0500320E

    Dialyzers were distributed after being exposed to cold temperatures which exceeded the low-temperature range in the firm's standard operating procedure.

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the state of Pennsylvania.Official notice →
All 33 Fresenius Medical Care Holdings recalls →

Other drugs & medical recalls in the United States

  1. 2022-02-16
  2. US
    Drugs & medicalClass II

    Avanos Medical, Inc. recalls MIC* Percutaneous Endoscopic Gastrostomy PEG Kit with ENFit Connectors - 20 Fr-PULL

    The External Retention Bolster contained in the kit may be a larger diameter than the tube.

    FDA · DeviceAvanos Medical, Inc.Worldwide distribution - US Nationwide and the countries of Canada and Chile.Official notice →
  3. US
    Drugs & medicalClass I

    E25Bio, Inc. recalls E25Bio COVID-19 Direct Antigen Rapid Test ; E25Bio COVID-19 DART Kit

    The Test is not authorized, cleared, or approved by FDA for clinical diagnostic or screening purposes. It was distributed without sufficient directions or controls in place.

    FDA · DeviceE25Bio, Inc.US Nationwide distribution in the states of MA, FL, and HI.Official notice →
  4. US
    Drugs & medicalClass II

    Argon Medical Devices, Inc recalls 1. Arterial Line Kit 20ga x 6" 2. Arterial Line Kit /1 20ga x 6"

    Guidewire included with two lots of Arterial Line Kits would not fit through the needle of the same kit. This could potentially lead to a slight delay in the procedure and minor blood loss.

    FDA · DeviceArgon Medical Devices, IncUS Nationwide distribution in the states of AL, CA, DE, IL, IN, LA, MT, NV,…Official notice →
  5. US
    Drugs & medicalClass II

    Gentueri Inc recalls Gentueri Viral Transport Medium Kit with 8cm Breakpoint Swab

    Kits failed sterility testing and showed evidence of contamination.

    FDA · DeviceGentueri IncDomestic: Illinois, Nebraska, and Wisconsin.Official notice →
  6. US
    Drugs & medicalClass I

    Esupplementsales, LLC recalls HARD DAWN, Rise and Shine capsules, 500 mg, packaged in 10-count blisters per carton, Made in Malaysia, UPC 680044 364926, ASIN: B077XCCL59

    Marketed Without an Approved NDA/ANDA: Firm was informed by Amazon that analytical testing showed the presence of tadalafil. Tadalafil is a FDA approved drug, making this product an unapproved drug.

    FDA · DrugEsupplementsales, LLCNationwide in the USA via Amazon at www.amazon.comOfficial notice →
  7. US
    Drugs & medicalClass I

    Junp Hydration LLC recalls MegMan Performance Booster, 800 mg capsules, packaged in 10-count blisters packaged in a carton, UPC 8 48998 00091 2, ASIN B08Z74KS88

    Marketed Without An Approved NDA/ANDA: product was found to contain undeclared tadalafil, an ingredient found in FDA approved products for the treatment of male sexual enhancement, making this an unapproved drug.

    FDA · DrugJunp Hydration LLCProduct was distributed nationwide in the USA via Amazon MarketplaceOfficial notice →
See all drugs & medical recalls →