Drugs & medicalClass II

Fresenius Medical Care Holdings, Inc. recalls AquaC UNO H [120V, 60Hz, 1700VA, Fresenius US-Version: 24-0970-1 Vivonic Part Number: G02000966-US

Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on…

Read the official notice →
Date
2022-08-03
Country
United States
Brand
Fresenius Medical Care Holdings
Product
AquaC UNO H [120V, 60Hz, 1700VA, Fresenius US-Version: 24-0970-1 Vivonic Part Number: G02000966-US
Classification
Class II
Units
US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Fresenius Medical Care Holdings recalls

  1. 2022-11-16
  2. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls NaturaLyte Liquid Acid concentrate, Part Number 08-0231-4, four (4) one-gallon bottles per bottle case. For dialysis.

    The "Nominal Chemical Composition" section of the label incorrectly lists 5.62g/L KCL (potassium chloride). 5.62g/L actually corresponds to the CaCl (calcium chloride) content. The product remains potassium free.

    FDA · DeviceFresenius Medical Care Holdings, Inc.Domestic distribution to AL AZ CA DC FL GA HI IL IN MD MI MO MS NE NJ NY OH PA…Official notice →
  3. 2022-10-26
  4. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls 0.9% Sodium Chloride Injection, USP, Each 100 mL contains: SODIUM CHLORIDE, USP - 900 mg, WATER FOR INJECTION, USP - qs,…

    Lack of Assurance of Sterility: Leakage of 0.9% Sodium Chloride for Injection, 1L, 12pk Saline Solution.

    FDA · DrugFresenius Medical Care Holdings, Inc.Nationwide in the USAOfficial notice →
  5. 2022-09-28
  6. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius…

    Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV,…Official notice →
  7. 2022-07-20
  8. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls NATURALYTE Liquid Acid 3.0K 3.0Ca- indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for…

    Alternate Bottle Caps labeled with "Sealed for your Protection" on the foil not qualified for use

    FDA · DeviceFresenius Medical Care Holdings, Inc.NatiowideOfficial notice →
  9. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ…

    AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device.…

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of NC and WV.Official notice →
  10. 2022-07-13
  11. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls NaturaLyte 2301 2K 3.0 Ca 1 Mg Gal, Model No. 08-2301-3

    The product was potentially exposed to below-recommended storage temperatures, which may cause leaks in the packaging. Leaks can lead to biological contamination and therefore an increased risk of potential bloodstream…

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of Texas and Louisiana.Official notice →
All 33 Fresenius Medical Care Holdings recalls →

Other drugs & medical recalls in the United States

  1. 2022-08-03
  2. US
    Drugs & medicalClass II

    Covidien Llc recalls TruClear Hysteroscopic Tissue Removal System Control Unit- designed to meet the requirements of intrauterine tissue removal. Model Number:…

    Electrical component failure within the control unit. The observed failure mode is a loss of function of the device when used with handpiece and foot switch.

    FDA · DeviceCovidien LlcAZ CA CO IL KS MA MI MN MO NY OK PA TN TX TX WA Foreign: United Kingdom…Official notice →
  3. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls Centricity PACS Software Version 7.0 SP0.0.4.7

    The Event Notification Manager (ENM) functionality for certain products does not process notifications for study modifications performed on post verified exams.

    FDA · DeviceGE Healthcare, LLCWorldwide distribution - US Nationwide distribution in the states of CO, NV,…Official notice →
  4. US
    Drugs & medicalClass III

    Zeiss, Carl Inc recalls ZEISS Axio Observer 5 microscope, Product material no. 431006-9901-000

    An incorrect Unique Device Identification (UDI) label was installed on some ZEISS Axio Observer 5 microscopes.

    FDA · DeviceZeiss, Carl IncUS Nationwide distribution in the state of NY.Official notice →
  5. US
    Drugs & medicalClass II

    NordicNeuroLab AS recalls nordicBrainEX, software versions 2.21 through 2.3.10

    When exporting merged results from the BOLD, DSC, and DCE modules of nordicBrainEx, where both the underlay and overlay originated from a multiframe dataset, the resulting output may have left/right sides flipped.

    FDA · DeviceNordicNeuroLab ASUS Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Touch US Llc recalls etouchus, model no. ETM-G01. Noninvasive Blood glucose meter.

    Device was marketed without FDA 510(k) clearance or a PMA.

    FDA · DeviceTouch US LlcUS Nationwide Distribution to: CA, FL IL KY LA MA MD MI NC NH NJ NY OH OK OR…Official notice →
  7. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls Da Vinci S/Si Large Clip Applier, Model 420230 Da Vinci S/Si Medium-Large Clip Applier, Model 420327 Da Vinci S/Si Small Clip Applier,…

    Due to the potential for unexpected motion caused by the partial disengagement of the instrument from the system.

    FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK,…Official notice →
See all drugs & medical recalls →