Drugs & medicalClass II

Fresenius Medical Care Holdings, Inc. recalls AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ…

AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device.…

Read the official notice →
Date
2022-07-20
Country
United States
Brand
Fresenius Medical Care Holdings
Product
AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA…
Classification
Class II
Units
US Nationwide distribution in the states of NC and WV.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Fresenius Medical Care Holdings recalls

  1. 2022-09-28
  2. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius…

    Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV,…Official notice →
  3. 2022-08-03
  4. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls AquaC UNO H [120V, 60Hz, 1700VA, Fresenius US-Version: 24-0970-1 Vivonic Part Number: G02000966-US

    Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on…

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution.Official notice →
  5. 2022-07-20
  6. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls NATURALYTE Liquid Acid 3.0K 3.0Ca- indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for…

    Alternate Bottle Caps labeled with "Sealed for your Protection" on the foil not qualified for use

    FDA · DeviceFresenius Medical Care Holdings, Inc.NatiowideOfficial notice →
  7. 2022-07-13
  8. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls NaturaLyte 2301 2K 3.0 Ca 1 Mg Gal, Model No. 08-2301-3

    The product was potentially exposed to below-recommended storage temperatures, which may cause leaks in the packaging. Leaks can lead to biological contamination and therefore an increased risk of potential bloodstream…

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of Texas and Louisiana.Official notice →
  9. 2022-07-06
  10. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls Optiflux 200NRe Dialyzer Finished Assy., Model no. 0500320E

    Dialyzers were distributed after being exposed to cold temperatures which exceeded the low-temperature range in the firm's standard operating procedure.

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the state of Pennsylvania.Official notice →
  11. 2022-02-16
  12. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls Novalung Sensor Box (F30000163) a component of Novalung Console (F30000162)- Indicated for long-term (> 6 hours)…

    Error messages 206 (yellow) and 208 (red) technical failure, flow measurement during use of the Novalung System imply that the communication between flow sensor and sensor box is interrupted. An interrupted…

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AL, AZ, CA, CO, FL, IL, MI, NJ,…Official notice →
All 33 Fresenius Medical Care Holdings recalls →

Other drugs & medical recalls in the United States

  1. 2022-07-20
  2. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS¿ ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS¿…

    On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU).

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Colombia, India,…Official notice →
  3. US
    Drugs & medicalClass II

    Wright Medical Technology, Inc. recalls Wright EVOLVE TRIAD Plate Cutter, REF 49510120

    The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use

    FDA · DeviceWright Medical Technology, Inc.Nationwide and PR, and Australia, Canada, Chile, Colombia, Hong Kong, Puerto…Official notice →
  4. US
    Drugs & medicalClass II

    Draeger Medical Systems, Inc. recalls Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system

    The firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of the system.

    FDA · DeviceDraeger Medical Systems, Inc.No US distribution. Foreign distribution to BY, CL, DE, ES, GB, HU, IN, IT,…Official notice →
  5. US
    Drugs & medicalClass II

    Cordis US Corp recalls PALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035" Delivery System, Catalog numbers PG2990BPS, PG2990BPX, PG3990BPS, PG3990BPX.

    Potential for stent dislodgement and associated failures related to two specific sizes of the device.

    FDA · DeviceCordis US CorpUS Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE,…Official notice →
  6. US
    Drugs & medicalClass II

    Cook Incorporated recalls Pressure Monitoring Set, Single-lumen uncoated central venous catheter. RPN C-PMS-250, GPN G02838

    The products may contain excess coating material on the outside and/or interior of the needle component. Potential adverse events that may occur if an affected product is used include a delay in the procedure, a…

    FDA · DeviceCook IncorporatedUS Nationwide. Brazil, Canada,Official notice →
  7. US
    Drugs & medicalClass II

    SCIENCE & BIO MATERIALS recalls ComposiTCP 30 Interference screw D 7- L:25 mm, REF 110004605

    Mislabeled product.

    FDA · DeviceSCIENCE & BIO MATERIALSUS Nationwide distribution.Official notice →