Fresenius Medical Care Holdings, Inc. recalls NaturaLyte 2301 2K 3.0 Ca 1 Mg Gal, Model No. 08-2301-3
The product was potentially exposed to below-recommended storage temperatures, which may cause leaks in the packaging. Leaks can lead to biological contamination and therefore an increased risk of potential bloodstream…
Read the official notice →- Date
- 2022-07-13
- Country
- United States
- Brand
- Fresenius Medical Care Holdings
- Product
- NaturaLyte 2301 2K 3.0 Ca 1 Mg Gal, Model No. 08-2301-3
- Classification
- Class II
- Units
- US Nationwide distribution in the states of Texas and Louisiana.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Fresenius Medical Care Holdings recalls
- 2022-09-28
US
Fresenius Medical Care Holdings, Inc. recalls AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius…
Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured
FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV,…Official notice →- 2022-08-03
US
Fresenius Medical Care Holdings, Inc. recalls AquaC UNO H [120V, 60Hz, 1700VA, Fresenius US-Version: 24-0970-1 Vivonic Part Number: G02000966-US
Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on…
- 2022-07-20
US
Fresenius Medical Care Holdings, Inc. recalls NATURALYTE Liquid Acid 3.0K 3.0Ca- indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for…
Alternate Bottle Caps labeled with "Sealed for your Protection" on the foil not qualified for use
US
Fresenius Medical Care Holdings, Inc. recalls AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ…
AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device.…
FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of NC and WV.Official notice →- 2022-07-06
US
Fresenius Medical Care Holdings, Inc. recalls Optiflux 200NRe Dialyzer Finished Assy., Model no. 0500320E
Dialyzers were distributed after being exposed to cold temperatures which exceeded the low-temperature range in the firm's standard operating procedure.
FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the state of Pennsylvania.Official notice →- 2022-02-16
US
Fresenius Medical Care Holdings, Inc. recalls Novalung Sensor Box (F30000163) a component of Novalung Console (F30000162)- Indicated for long-term (> 6 hours)…
Error messages 206 (yellow) and 208 (red) technical failure, flow measurement during use of the Novalung System imply that the communication between flow sensor and sensor box is interrupted. An interrupted…
FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AL, AZ, CA, CO, FL, IL, MI, NJ,…Official notice →
Other drugs & medical recalls in the United States
- 2022-07-13
US
Covidien, LP recalls EEA Autosuture Circular Stapler with DST Series Technology, 25mm. Model Numbers EEA25, EEAXL25, EEA2535, and EEAXL2535.
The affected staplers have the potential for the staple guide to not be securely attached to the instrument. The issue could cause the component to disengage and if disengaged, could allow the device to transect tissue…
US
Radiometer Medical ApS recalls ABL800 Flex Analyzer Model Numbers 393-800 and 393-801
There is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+.
FDA · DeviceRadiometer Medical ApSWorldwide distribution - US Nationwide distribution in the states of AR, AZ,…Official notice →US
Covidien Llc recalls Palindrome Precision Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis,…
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
FDA · DeviceCovidien LlcWorldwide distribution - US Nationwide distribution in the states of AK, AL,…Official notice →US
American Contract Systems, Inc. recalls Clean Catch Urine Kit NS 178489, AHMU58A
Affected products were assembled off-site by individuals who may not have been properly trained.
FDA · DeviceAmerican Contract Systems, Inc.US distribution in the state of Minnesota.Official notice →US
Wedgewood Village Pharmacy, LLC recalls Bethanechol Chloride 1mg/ml, 80 ml Suspension bottles Banana/Strawberry, Qty:1, Wedgewood Village Pharmacy LLC, 405 Heron Dr.…
Lack of Processing Controls: Insanitary conditions observed during recent FDA Inspection
FDA · DrugWedgewood Village Pharmacy, LLCProducts were distributed nationwide in the USA.Official notice →US
Golden State Medical Supply Inc. recalls Enalapril Maleate Tablets, USP; 2.5 mg; 90 tablets, NDC 60429-183-90; manufactured by Taro Pharmaceutical Industries, Ltd.;…
CGMP Deviations
FDA · DrugGolden State Medical Supply Inc.Nationwide in the U.S: PA, MA, OH, TX, IL, ND, HIOfficial notice →