Drugs & medicalClass II

Fresenius Medical Care Holdings, Inc. recalls NaturaLyte Liquid Acid concentrate, Part Number 08-0231-4, four (4) one-gallon bottles per bottle case. For dialysis.

The "Nominal Chemical Composition" section of the label incorrectly lists 5.62g/L KCL (potassium chloride). 5.62g/L actually corresponds to the CaCl (calcium chloride) content. The product remains potassium free.

Read the official notice →
Date
2022-11-16
Country
United States
Brand
Fresenius Medical Care Holdings
Product
NaturaLyte Liquid Acid concentrate, Part Number 08-0231-4, four (4) one-gallon bottles per bottle case. For dialysis.
Classification
Class II
Units
Domestic distribution to AL AZ CA DC FL GA HI IL IN MD MI MO MS NE NJ NY OH PA…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Fresenius Medical Care Holdings recalls

  1. 2023-06-07
  2. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls Sodium Chloride 0.9% Injection, USP, 1000 mL bags, 12 bags per carton, Rx only, Fresenius Medical Care North America,…

    Lack of assurance of sterility: The product was potentially exposed to below-recommended storage temperatures, which may cause leaks in the packaging.

    FDA · DrugFresenius Medical Care Holdings, Inc.Product was distributed to 32 clinics in SC, NC and VA.Official notice →
  3. 2022-12-28
  4. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls 2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare…

    (1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false…

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution.Official notice →
  5. 2022-12-14
  6. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls 0.9% Sodium Chloride Injection, USP, Each 100 mL contains: SODIUM CHLORIDE, USP - 900 mg, WATER FOR INJECTION, USP - qs,…

    Lack of Assurance of Sterility: Leakage of 0.9% Sodium Chloride for Injection, 1L, 12pk Saline Solution.

    FDA · DrugFresenius Medical Care Holdings, Inc.Product was distributed Nationwide in the USA.Official notice →
  7. 2022-10-26
  8. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls 0.9% Sodium Chloride Injection, USP, Each 100 mL contains: SODIUM CHLORIDE, USP - 900 mg, WATER FOR INJECTION, USP - qs,…

    Lack of Assurance of Sterility: Leakage of 0.9% Sodium Chloride for Injection, 1L, 12pk Saline Solution.

    FDA · DrugFresenius Medical Care Holdings, Inc.Nationwide in the USAOfficial notice →
  9. 2022-09-28
  10. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius…

    Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV,…Official notice →
  11. 2022-08-03
  12. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls AquaC UNO H [120V, 60Hz, 1700VA, Fresenius US-Version: 24-0970-1 Vivonic Part Number: G02000966-US

    Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on…

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution.Official notice →
All 33 Fresenius Medical Care Holdings recalls →

Other drugs & medical recalls in the United States

  1. 2022-11-16
  2. US
    Drugs & medicalClass II

    Ferno-Washington Inc recalls FERNO POWER X1 AMBULANCE COT- Item #0015807/UDI-DI 00190790002183 with 36V Li-Ion Batteries Model Number #0822484/0822483 Note: Transcend…

    Batteries could become unstable when not properly maintained and/or subjected to repeated drops, power washing directly into battery contacts or abuse, may result in a fire and/or injury

    FDA · DeviceFerno-Washington IncNationwide Foreign: BRAZIL BRUNEI CHILE CHINA PERU POLAND PUERTO RICO…Official notice →
  3. US
    Drugs & medicalClass II

    iRhythm Technologies, Inc. recalls Zio AT Clinical Reference Manual, Part: ALB0031, and Zio AT Important Information Pamphlet, Part: ALB0034. For Zio Ambulatory ECG…

    Ambulatory ECG monitoring system labeling update: 1) Clinically actionable arrhythmia notifications delayed until patient registration is complete; 2) Patches limited to 100 symptomatic and 500 asymptomatic wireless…

    FDA · DeviceiRhythm Technologies, Inc.US: MD, NY, FL, TN, WA, AL, MA, AZ, CA, KS, HI, PA, NJ, TX, MN, VA, AR, OR,…Official notice →
  4. US
    Drugs & medicalClass II

    SEDECAL SA recalls Sedecal Soltus Battery Mobile X-Ray System REF Soltus 450 Battery Mobile X-Ray System REF Soltus 500

    Due to un-commanded movement of mobile x-ray unit.

    FDA · DeviceSEDECAL SAU.S.: KY, WI, FL, DE, WV, CA, MO, NY, NC, AL, AR, MS, GA, IL, IA, TN, KS, KY,…Official notice →
  5. US
    Drugs & medicalClass I

    Viatris Inc recalls Octreotide Acetate Injection 500 mcg/mL, 10 x 1 mL Single-Dose Unit-of-Use Syringes, For Subcutaneous or Intravenous Use, Rx Only,…

    Presence of Particulate Matter: Product complaint for the presence of glass particles in a syringe.

    FDA · DrugViatris IncNationwide in the USAOfficial notice →
  6. 2022-11-09
  7. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Simple Clamp, US/EURO Rails, UPPER EXTREMITY DISTRACTION CMPT, REF F-40018, #800101. To position, support and/or distract the…

    the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.

    FDA · DeviceBaxter Healthcare CorporationMI, BA, WI, WA, TX, and Germany, Japan, South Korea, Sweden, South AfricaOfficial notice →
  8. US
    Drugs & medicalClass II

    Becton Dickinson Infusion Therapy Systems Inc. recalls BD Nexiva Closed IV Catheter System - Single Port REF 383516 20 GA X1.0o in (1.1 x 25 mm) 61 mL/min (3660 mL/hr)

    During needle withdrawal, the tip shield may prematurely separate from the catheter adapter and preventing the needle tip from being in the safely shielded (leaving the needle tip exposed).

    FDA · DeviceBecton Dickinson Infusion Therapy Systems Inc.U.S.: New York, Florida, Maryland, North Carolina, Virginia, Georgia, Rhode…Official notice →