Drugs & medicalClass II

Draeger Medical, Inc. recalls Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up…

Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds

Read the official notice →
Date
2021-10-20
Country
United States
Brand
Draeger Medical
Product
Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up…
Classification
Class II
Units
US Nationwide
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Draeger Medical recalls

  1. 2023-07-19
  2. US
    Drugs & medicalClass I

    Draeger Medical, Inc. recalls Oxylog 3000 Plus emergency and transport ventilator

    Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.

    FDA · DeviceDraeger Medical, Inc.Distributed Nationwide. OUS Distribution: Afghanistan, Algeria, Andorra,…Official notice →
  3. 2023-05-24
  4. US
    Drugs & medicalClass I

    Draeger Medical, Inc. recalls VentStar Basic (P)180, MP00351

    Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.

    FDA · DeviceDraeger Medical, Inc.Domestic: US Nationwide Distribution.Official notice →
  5. 2023-03-01
  6. US
    Drugs & medicalClass I

    Draeger Medical, Inc. recalls NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size L)

    The magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.

    FDA · DeviceDraeger Medical, Inc.US Nationwide DistributionOfficial notice →
  7. 2022-06-22
  8. US
    Drugs & medicalClass I

    Draeger Medical, Inc. recalls Drager SafeStar 55, Catalog No. MP01790

    Possible occlusion of filters due to manufacturing error.

    FDA · DeviceDraeger Medical, Inc.Worldwide distribution - US Nationwide distribution in the states of AZ, CA,…Official notice →
  9. 2022-04-13
  10. US
    Drugs & medicalClass II

    Draeger Medical, Inc. recalls Evita V500 Ventilator with Software 2.51.01 and Lower - intended for the ventilation of adult, pediatric and neonatal patients Catalog…

    Cybersecurity Improvement Action-not equipped against potential cyber security threats

    FDA · DeviceDraeger Medical, Inc.US NationwideOfficial notice →
  11. 2022-01-26
  12. US
    Drugs & medicalClass II

    Draeger Medical, Inc. recalls Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300

    26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas…

    FDA · DeviceDraeger Medical, Inc.US Nationwide distribution in the states of NY, PA, and WI.Official notice →
All 11 Draeger Medical recalls →

Other drugs & medical recalls in the United States

  1. 2021-10-20
  2. US
    Drugs & medicalClass II

    ELEKTA SOLUTIONS AB recalls Elekta Medical Linear Accelerator (EMLA): Elekta Harmony Pro, REF 1555143; Elekta Infinity, REF 1552091, and Elekta Versa HD, REF 1552093.

    There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

    FDA · DeviceELEKTA SOLUTIONS ABDistribution was to ND and NC. There was no military/government distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Flower Orthopedics Corporation recalls Flower Drill Bit Kit 2.0 mm - intended to be used with solid 2.0mm screws and can be used with any plate that has been cleared as…

    Drill bits packaged in kits are missing the AO connect feature and may result in delay of procedure

    FDA · DeviceFlower Orthopedics CorporationArizona California Colorado Florida Florida Illinois Indiana Nevada Ohio Texas…Official notice →
  4. US
    Drugs & medicalClass II

    Uoc Usa Inc recalls United Orthopedics USTAR II Knee System USTAR II Knee System Segment Part, RHS, 25 mm Length, Catalog # 2915-1025

    Due to product not meeting specification for concentricity and failing to meet requirements of dynamic fatigue testing.

    FDA · DeviceUoc Usa IncU.S. Nationwide: CA, CO, and NV No O.U.S.Official notice →
  5. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls Ion Flexison Biopsy Needle, 19G, Extension Range : 0-3 cm, Ref: 490104-09 and 49004-10, 5-pack, Sterile EO, Rx only, UDI:00886874115978

    Potential for the sheath tip on the biopsy needle to separate from the sheath shaft.

    FDA · DeviceIntuitive Surgical, Inc.US: AZ, AR, CA, CO, DC, FL, GA, IL, IN, IO, MD, MA, MI, MN, MO, NV, NJ, NY,…Official notice →
  6. US
    Drugs & medicalClass II

    Centese Inc. recalls Thoraguard Chest Tube Kit, 20 Fr

    An adhesive joint in the SmartValve component of the device can become compromised resulting in a leak.

    FDA · DeviceCentese Inc.Distribution to a single consignee located in California.Official notice →
  7. US
    Drugs & medicalClass II

    K2M, Inc recalls Chesapeake Anterior Lumbar (AL) Removal Tool; Cat. no. 2008-90068

    Units may arrive with missing pins and/or in a state in disassembly. Pins dislodging from the instrument during a procedure may enter the surgical cavity, requiring intraoperative removal or revision surgery.

    FDA · DeviceK2M, IncDomestic distribution to CO, IL, LA, MA, MS, OK, PA, TN, and TX. Foreign…Official notice →
See all drugs & medical recalls →