Flower Orthopedics Corporation recalls Flower Drill Bit Kit 2.0 mm - intended to be used with solid 2.0mm screws and can be used with any plate that has been cleared as…
Drill bits packaged in kits are missing the AO connect feature and may result in delay of procedure
Read the official notice →- Date
- 2021-10-20
- Country
- United States
- Brand
- Flower Orthopedics
- Product
- Flower Drill Bit Kit 2.0 mm - intended to be used with solid 2.0mm screws and can be used with any plate that has been cleared as part of…
- Classification
- Class II
- Units
- Arizona California Colorado Florida Florida Illinois Indiana Nevada Ohio Texas…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Flower Orthopedics recalls
- 2023-10-18
US
Flower Orthopedics Corporation recalls Coventus Flower Orthopedics Flex-Thread TM 3.2mm Cannulated Reamer-used to ream and subsequently open the intramedullary canal to…
Device is breaking while in use, potential for patients to require a revision surgery.
- 2022-01-12
US
Flower Orthopedics Corporation recalls Flex-Thread Inserter Shaft intended for Fibula intramedullary-intended to be used with the Inserter Draw Rod and Blue Ratchet…
Mating incompatibility between the Inserter Shaft and the Inserter Draw Rod causing the inserter draw rod to bind and not rotate freely as intended surgeon may be unable to complete surgery unless a second set is…
FDA · DeviceFlower Orthopedics CorporationUS Nationwide distribution in the states of AZ, CA, CO, GA, TN, FL, IL, MA,…Official notice →- 2021-10-13
US
Flower Orthopedics Corporation recalls Flower E-Kit, Advanced-Bone Fixation Fasterner-Fixation of Ref: EWK 201
Product kit missing CDG 200 (Cannulated Countersink for 3.0-4.5mm Screws) from EWK 201 (Flower E-Kit, Advanced)
FDA · DeviceFlower Orthopedics CorporationUS Nationwide distribution in the states of AZ, CA, CO,CT, FL, IL, IN, NC, NY,…Official notice →
Other drugs & medical recalls in the United States
- 2021-10-20
US
Beckman Coulter Inc. recalls Power Processor 3K, 3K High Speed Stockyards Module Nameplate: PP 3K Stockyard (Refrig) 9707 REF A88116
Missing splash guard which may expose users to hazardous materials. .
FDA · DeviceBeckman Coulter Inc.U.S. Nationwide Distribution in the states of: CA, GA, IA, ID, IL, IN, LA, MA,…Official notice →US
Elekta, Inc. recalls Elekta Medical Linear Accelerator (EMLA), configured as follows: a. Elekta Precise Digital Accelerator; b. Elekta Harmony; c. Elekta…
There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.
FDA · DeviceElekta, Inc.Distributed nationwide to FL, NC, OH, CA, IL, MD, OR, AL, NY, PA, MA, AK, GA,…Official notice →US
ELEKTA SOLUTIONS AB recalls Elekta Medical Linear Accelerator (EMLA): Elekta Harmony Pro, REF 1555143; Elekta Infinity, REF 1552091, and Elekta Versa HD, REF 1552093.
There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.
FDA · DeviceELEKTA SOLUTIONS ABDistribution was to ND and NC. There was no military/government distribution.Official notice →US
Uoc Usa Inc recalls United Orthopedics USTAR II Knee System USTAR II Knee System Segment Part, RHS, 25 mm Length, Catalog # 2915-1025
Due to product not meeting specification for concentricity and failing to meet requirements of dynamic fatigue testing.
US
Draeger Medical, Inc. recalls Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up…
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds
US
Intuitive Surgical, Inc. recalls Ion Flexison Biopsy Needle, 19G, Extension Range : 0-3 cm, Ref: 490104-09 and 49004-10, 5-pack, Sterile EO, Rx only, UDI:00886874115978
Potential for the sheath tip on the biopsy needle to separate from the sheath shaft.
FDA · DeviceIntuitive Surgical, Inc.US: AZ, AR, CA, CO, DC, FL, GA, IL, IN, IO, MD, MA, MI, MN, MO, NV, NJ, NY,…Official notice →