Draeger Medical, Inc. recalls Oxylog 3000 Plus emergency and transport ventilator
Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.
Read the official notice →- Date
- 2023-07-19
- Country
- United States
- Brand
- Draeger Medical
- Product
- Oxylog 3000 Plus emergency and transport ventilator
- Classification
- Class I
- Units
- Distributed Nationwide. OUS Distribution: Afghanistan, Algeria, Andorra,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Draeger Medical recalls
- 2024-11-20
US
Draeger Medical, Inc. recalls Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable Component: Not applicable
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
FDA · DeviceDraeger Medical, Inc.US Nationwide distribution in the states of AL, AZ, CA, FL, GA, IA, IL, KY,…Official notice →- 2024-04-24
US
Draeger Medical, Inc. recalls Dr¿ger Perseus A500 Anesthesia Workstation
Unexpected shutdown while operating on battery power.
FDA · DeviceDraeger Medical, Inc.Domestic: Nationwide Distribution; Foreign: Angola, Argentina, Australia,…Official notice →- 2023-08-30
US
Draeger Medical, Inc. recalls Carina Sub-Acute Care Ventilator
Presence of polyether polyurethane (PE-PUR) and emission of 1,3-Dichloropropan-2-ol that exceed the acceptable uptake level during continuous use (>30 days) in pediatric patients.
- 2023-05-24
US
Draeger Medical, Inc. recalls VentStar Basic (P)180, MP00351
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- 2023-03-01
US
Draeger Medical, Inc. recalls NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size L)
The magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.
- 2022-06-22
US
Draeger Medical, Inc. recalls Drager SafeStar 55, Catalog No. MP01790
Possible occlusion of filters due to manufacturing error.
FDA · DeviceDraeger Medical, Inc.Worldwide distribution - US Nationwide distribution in the states of AZ, CA,…Official notice →
Other drugs & medical recalls in the United States
- 2023-07-19
US
Baxter Healthcare Corporation recalls Allen Advance Chest Support with Pad
Potential for the device to crack where the Chest Base Prone support attaches to the carbon fiber operating room (OR) table rail.
FDA · DeviceBaxter Healthcare CorporationWorldwide distribution - US Nationwide and the countries of Canada, Australia,…Official notice →US
Osteomed, LLC recalls REF: 218-1604, 1.6mm x 4mm Profile Zero Auto-Drive Screw, Rx Only, Sterile
Screws for osteotomies have contain an undersize screwhead which has the potential to pass through the plate hole.
FDA · DeviceOsteomed, LLCWorldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, KS,…Official notice →US
Baxter Healthcare Corporation recalls The Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009
A medical device correction is being issued for SIGMA Spectrum Infusion System (V8 Platform) and Spectrum IQ Infusion System with Dose IQ Safety Software infusion pumps due to an increase in reported false upstream…
US
Cipla USA, Inc. recalls Albuterol Sulfate Inhalation Aerosol, 90 mcg, 200 Metered Inhalation, net content 6.7 g canister packaged in a box, Rx only, Manufactured…
Defective container: empty inhaler and leakage observed through the inhaler valve due to partially missing bottom seat (gasket).
US
Global Pharma Healthcare Private Limited recalls Artificial Tears (Carboxymethylcellulose Sodium) Lubricant Eye Drops, 10 MG in 1 ml, 1/2 fl oz (15 ml) bottle,…
CGMP Deviations: All other lots of eye drops are being recalled due to CGMP Deviations because they were manufactured in the same facility under the same conditions as the lots found to be contaminated.
FDA · DrugGlobal Pharma Healthcare Private LimitedProduct was distributed to two distributors who further distributed Nationwide…Official notice →US
Global Pharma Healthcare Private Limited recalls Artificial Tears (Carboxymethylcellulose Sodium) Lubricant Eye Drops, 10 MG in 1 ml, 1/2 fl oz (15 ml) bottle,…
Non-Sterility: FDA analysis found unopened products to have bacterial contamination.
FDA · DrugGlobal Pharma Healthcare Private LimitedProduct was distributed to two distributors who further distributed Nationwide…Official notice →