K2M, Inc recalls Chesapeake Anterior Lumbar (AL) Removal Tool; Cat. no. 2008-90068
Units may arrive with missing pins and/or in a state in disassembly. Pins dislodging from the instrument during a procedure may enter the surgical cavity, requiring intraoperative removal or revision surgery.
Read the official notice →- Date
- 2021-10-20
- Country
- United States
- Brand
- K2M
- Product
- Chesapeake Anterior Lumbar (AL) Removal Tool; Cat. no. 2008-90068
- Classification
- Class II
- Units
- Domestic distribution to CO, IL, LA, MA, MS, OK, PA, TN, and TX. Foreign…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More K2M recalls
- 2022-07-06
US
K2M, Inc. recalls Yukon Straight Rod 4.0 x120mm; Cat# 7601-540120. Used to provide immobilization and stabilization of spinal segments.
Some of the rods in this lot were observed to have discoloration, later confirmed to be indication of trace amounts of inorganic phosphorous compound. This may produce a mild irritating or inflammatory response;…
FDA · DeviceK2M, Inc.Domestic distribution to FL GA NE NV NY OH OK SC Foreign distribution to…Official notice →- 2022-03-09
US
K2M, Inc recalls Everest MI XT Inner Dilator, Catalog Number 5101-90167
Stryker received two (2) complaints for units from specific lots of Inner and Outer Dilators not fitting together properly during a surgical procedure. Some dilators from these lots and from other lots were…
Other drugs & medical recalls in the United States
- 2021-10-20
US
Draeger Medical, Inc. recalls Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up…
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds
US
Intuitive Surgical, Inc. recalls Ion Flexison Biopsy Needle, 19G, Extension Range : 0-3 cm, Ref: 490104-09 and 49004-10, 5-pack, Sterile EO, Rx only, UDI:00886874115978
Potential for the sheath tip on the biopsy needle to separate from the sheath shaft.
FDA · DeviceIntuitive Surgical, Inc.US: AZ, AR, CA, CO, DC, FL, GA, IL, IN, IO, MD, MA, MI, MN, MO, NV, NJ, NY,…Official notice →US
Centese Inc. recalls Thoraguard Chest Tube Kit, 20 Fr
An adhesive joint in the SmartValve component of the device can become compromised resulting in a leak.
US
Smiths Medical ASD Inc. recalls smiths medical medex CBSS-5ML, Triple set, (172/15cm), Model Number DPSHC0092
Reported pressure shifts and incorrect pressure readings in LogiCal Closed Blood Sampling System sets.
FDA · DeviceSmiths Medical ASD Inc.Austria, France, Germany, Great Britain, and Saudi ArabiaOfficial notice →US
Red Medical Supplies recalls EcoGel 200 - RED Medical Supplies Blue Ultrasound Gel, Model No. 250UG-BMU
Ultrasound gels and lotions manufactured by Eco-Med Pharmaceutical, Inc. may contain bacterial contamination. On August 18, 2021,FDA published a Letter to Health Care Providers to further address this issue:…
FDA · DeviceRed Medical SuppliesFirm sold product through Amazon.com. Domestic distribution nationwide.Official notice →US
Inpeco S.A. recalls FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2). (Different brand names of same product.)
The Firmware (FW) of the Automation System Interface Module to ADVIA Centaur XP/XPT may lead to an incorrect association of test results to sample ID. Improper management of the sample may bring an incorrect patient…