Drugs & medicalClass I

Datex-Ohmeda, Inc. recalls EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002

GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These…

Read the official notice →
Date
2024-02-07
Country
United States
Brand
Datex-Ohmeda
Product
EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002
Classification
Class I
Units
Worldwide
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Datex-Ohmeda recalls

  1. 2026-09-23
  2. US
    Drugs & medicalClass II

    Datex-Ohmeda, Inc. recalls Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

    Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling…

    FDA · DeviceDatex-Ohmeda, Inc.Worldwide - US Nationwide distribution.Official notice →
  3. 2026-06-24
  4. US
    Drugs & medicalClass I

    Datex-Ohmeda Inc recalls Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001,…

    Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting…

    FDA · DeviceDatex-Ohmeda IncPendingOfficial notice →
  5. 2024-11-06
  6. US
    Drugs & medicalClass I

    DATEX--OHMEDA, INC. recalls GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3)…

    Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week.…

    FDA · DeviceDATEX--OHMEDA, INC.worldwideOfficial notice →
  7. 2024-10-30
  8. US
    Drugs & medicalClass I

    Datex-Ohmeda, Inc. recalls Aisys CS2 with Et Control. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to…

    Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

    FDA · DeviceDatex-Ohmeda, Inc.Worldwide distribution - US Nationwide and the countries of Aland Islands,…Official notice →
  9. 2024-10-23
  10. US
    Drugs & medicalClass I

    DATEX--OHMEDA, INC. recalls GE Heathcare Giraffe OmniBed; neo natal incubator

    GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This…

    FDA · DeviceDATEX--OHMEDA, INC.Worldwide distribution.Official notice →
  11. 2023-07-12
  12. US
    Drugs & medicalClass II

    DATEX--OHMEDA, INC. recalls Giraffe OmniBed Carestation CS1, Model Numbers: a. 2082844-001-XXX, b. 2074920-001-XXXXXX, c. M1050205, d. M1135822, e. M1145475, f.…

    GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe…

    FDA · DeviceDATEX--OHMEDA, INC.Worldwide distribution.Official notice →
All 7 Datex-Ohmeda recalls →

Other drugs & medical recalls in the United States

  1. 2024-02-07
  2. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion system, Interventional fluoroscopic x-ray system. System Product Name - Model Numbers Azurion 3M12…

    System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the…

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Ethicon Endo-Surgery Inc recalls ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, Endoscopic Linear Cutter Reload…

    There is the potential for incomplete staple line formation during the firing sequence, without most or all staples, which may lead to insufficient tissue apposition that could require surgical intervention. This may…

    FDA · DeviceEthicon Endo-Surgery IncWorldwide distribution - US Nationwide and the countries of Canada and Brazil.Official notice →
  4. US
    Drugs & medicalClass II

    American Contract Systems, Inc. recalls Procedural convenience kits and trays, cardiac, labeled as: a) EP LAB BI V IMPLANT PACK, kit number AGBI64J; b) LAB DEVICE…

    During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked"…

    FDA · DeviceAmerican Contract Systems, Inc.US Nationwide distribution.Official notice →
  5. US
  6. US
    Drugs & medicalClass II

    Den-Mat Holdings, LLC recalls Hydrogen Peroxide Oral Rinse, Significantly Reduces Bacteria, Fresh Mint Flavor, Alcohol Free, a)16 fl. oz. (473 mL) bottle, NDC…

    CGMP Deviations: products may not conform to the labeled specifications.

    FDA · DrugDen-Mat Holdings, LLCNationwide in the USOfficial notice →
  7. US
    Drugs & medicalClass II

    Amerisource Health Services LLC recalls Rifampin Capsules USP, 150 mg, 30 Capsules (3 x 10) unit doses per carton, Rx Only, Distributed by: American Health Packaging,…

    Failed Impurities/Degradation Specification.

    FDA · DrugAmerisource Health Services LLCNationwide in the USAOfficial notice →
See all drugs & medical recalls →